Manual

26
D24042 | Date 09/2014
AtmoCONTROL
4.4.3
Programme example
sterilisation
1 2 3
5
4
Tolerance band
6
180 °C
20 °C
3:00 h
1
2
3
6
4
7
0 % 100 % 0 %
100 % 0 %
Fig. 10
At the beginning, the fan is switched on to 100 % (1) and the air flap is closed (0 %) (2). Then,
Fig. 10
At the beginning, the fan is switched on to 100 % (1) and the air flap is closed (0 %) (2). Then,
Fig. 10
the appliance heats up to 180.0
At the beginning, the fan is switched on to 100 % (1) and the air flap is closed (0 %) (2). Then,
the appliance heats up to 180.0
At the beginning, the fan is switched on to 100 % (1) and the air flap is closed (0 %) (2). Then,
°
C (3) and maintains this temperature for 3 hours (4). The set-
At the beginning, the fan is switched on to 100 % (1) and the air flap is closed (0 %) (2). Then,
C (3) and maintains this temperature for 3 hours (4). The set-
At the beginning, the fan is switched on to 100 % (1) and the air flap is closed (0 %) (2). Then,
ting „Safe“ (5) ensures that the sterilisation time does not start (6) before the set tolerance band
the appliance heats up to 180.0
ting „Safe“ (5) ensures that the sterilisation time does not start (6) before the set tolerance band
the appliance heats up to 180.0
C (3) and maintains this temperature for 3 hours (4). The set-
ting „Safe“ (5) ensures that the sterilisation time does not start (6) before the set tolerance band
C (3) and maintains this temperature for 3 hours (4). The set-
(7) is reached and is restarted if it is exceeded.
ting „Safe“ (5) ensures that the sterilisation time does not start (6) before the set tolerance band
(7) is reached and is restarted if it is exceeded.
ting „Safe“ (5) ensures that the sterilisation time does not start (6) before the set tolerance band