Delivery, Transport and Setting Up
3.4
Setting up
The cooled incubator may only be placed on the ground, not on a table. When doing this,
ensure that the appliance is positioned exactly horizontally
.
The installation site must be level
and must be able to reliably carry the weight of the cooled incubator (see page
17
). Do not
place the appliance on inflammable surfaces and do not stack.
The front swivel castors can be locked with a catch. To ensure stability, always turn the front
swivel castors to the front to lock them.
A
power connection corresponding to the information on the nameplate (see page
17
) must
be available at the installation site.
The distance between the wall and the rear of the cooled incubator must be at least 15 cm.
The clearance from the ceiling must not be less than 20 cm and the side clearance from the
wall must not be less than 8 cm (
Fig. 8
). The ventilation slits of the
cooling unit
on the rear of
the appliance may not be adjusted.
Sufficient air circulation in the vicinity of the appliance must be guaranteed at all times.
≥ 15 cm
≥ 8 cm
≥ 8 cm
≥ 20 cm
Fig. 8
Minimum clearance from walls and ceiling