Warning Messages and Malfunctions
6.
Warning Messages and
Malfunctions
6.1
Warning messages
The warning messages also set off an intermittent acoustic signal:
This can be temporarily switched off by pressing the SET key.
Error in the temperature control system (see also chapter "
Error in the temperature control system (see also chapter "
Temperature monitoring
Temperature monitoring
"
Temperature monitoring " Temperature monitoring
on page
on page
53
):
):
53 ):53
Tb active -
if the
Tb active - if the Tb active -
temperature limiter is
triggered
Remedy: See page
54
hi-alarm -
if
hi-alarm - if hi-alarm -
overtemperature
protection is triggered
Remedy: Check the setting of the
MAX temperature monitor.
lo alarm -
if
undertemperature
protection is triggered
Remedy: Check the setting of the
MIN temperature monitor.
asf-alarm -
if the
automatic monitoring
function is triggered
Remedy: Check the setting of the
ASF temperature monitor.
Error in the dehumidification system:
Error in the dehumidification system:
if the water supply is
defective
Remedy: Check that the water
supply is properly connected. Fill
water supply tank with distilled
water if it is empty.
If humidity exceeds the
setpoint for longer than
30 minutes.
Remedy: Open the door for
30 sec. and wait to see if the
controller steadily adjusts to
the setpoint. If the error occurs
again, contact the customer
service.