Operating instructions
PERFECT page 14
Setting example “Normal operation“
The oven  (UFP500)  has to  heat  up  to  180°C  at  50%  fan  speed  and  with the  air  valve 20%  open.  The 
monitor function has to operate at 200°C.
180°C
monitor temperature 200°C
°C
h
1. Select operating mode “Normal operation“
After holding down the SET key (approx. 3 sec), the current operating 
mode is flashing. Select operating mode I with the push/turn control 
while holding down the SET key.
After the SET key has been released the controller is in 
operating mode I.
PRINT
SETUP
2. Select temperature setpoint
Hold  down  the  SET  key  and  use  the  push/turn  control  to  select  the 
required temperature setpoint of 180 °C.
After  the  SET  key  has  been released  the  oven  briefly  flashes  the 
temperature setpoint. The display then changes to the actual temperature 
and the controller starts to control to the selected temperature setpoint 
180 °C.
Heating is indicated by the orange heater symbol. 
˚C
3. Select fan speed
Turn the push/turn control clockwise until the fan symbol is flashing.
While holding down the SET key, use the push/turn control to set 50% 
fan speed.
4. Set air valve for air vent
Turn the push/turn control clockwise until the air valve symbol is 
flashing.
While holding down the SET key, use the push/turn control to set the 
air valve to 20%.
5. Select monitor temperature
Turn the push/turn control clockwise until the overtemperature display  
MAX
is flashing. Hold down the SET key and use the push/turn control 
to set the monitor temperature to 200 °C.
MIN
AUTO
MAX
°C