Operating instructions
PERFECT page 22
14. Set closure command for second ramp segment
Turn the push/turn control clockwise until a segment closure command 
(e.g end ) appears.
Hold down the SET key and set next with the push/turn control.
15. Select duration of third ramp segment
Using the push/turn control select the time indication
Hold  down  the  SET  key  and  set  the  time  08:00  using  the  push/turn  
control.
h
16. Select temperature of third ramp segment
Turn the  push/turn  control  clockwise  until  the  temperature  display  is 
flashing.
Hold down the SET key and set the required temperature setpoint of 
60 °C using the push/turn control.
˚C
17. Select fan speed for third ramp segment
Turn the push/turn control clockwise until the fan symbol is flashing.
Hold down the SET key and set the fan speed to 50% using the push/
turn control.
18. Set air valve for third ramp segment
Turn the push/turn control clockwise until the air valve symbol is 
flashing.
Hold down the SET key and set the air valve to 20% using the push/turn 
control
19. Set closure command for third ramp segment
Turn the push/turn control clockwise until a segment closure command 
(e.g end ) appears.
Press the SET key briefly to enter.
20. Exit programme writing mode EDIT
Turn the push/turn control clockwise until EXIT appears on the display. 
Press the SET key briefly to enter.
21. Set temperature monitor
Turn the push/turn control clockwise and set the temperature monitor. 
(see Section “Temperature monitor”)
MIN
MAX
AUTO
°C
22. Start programme
Turn the  push/turn  control  anticlockwise  until  the  stop  symbol   is 
flashing.
Hold down the SET key and select Start   with the push/turn control.