13  Temperature monitor and protection devices
The monitor temperature is measured with a separate PT100 temperature sensor inside the chamber. The 
monitor unit provides protection for the oven load as well as protection for oven and its surroundings.
The oven is provided with duplicate overtemperature protection (mechanical / electronic) according to DIN 12 880.
MIN
MAX
AUTO
°C
13.1  Mechanical temperature monitor: temperature limiter (TB)
All ovens of the PERFECT series are equipped with a mechanical temperature limiter (TB) Protection Class1 
to DIN 12 880.
If the electronic monitor system should fail during operation and the fixed factory-set maximum temperature 
is  exceeded  by  approx.  20°C  the  temperature  limiter  switches  off  the  heating  permanently  as  a  final 
protective measure. The alarm symbol lights up as warning 
Fault rectification after the TB cut-out has been activated:
Switch off the oven and allow it to cool down1. 
Rectify the fault (e.g. replace temperature probe) and where appropriate contact customer service2. 
The oven is again ready for operation only after it has cooled down and after the fault has been 3. 
  rectified
Overtemperature 
protection
 (TWW, TWB)
Undertemperature protection
Audible 
alarm signal
Automatic 
temperature monitor 
(ASF)
visual alarm symbol
alight:  TB alarm
flashing:  TWW alarm   
      TWB-alarm
      ASF-alarm