13.2 Electronic temperature monitor
13.2.1 Overtemperature protection 
MAX
Range: up to 10°C max above nominal 
temperature (for nominal temperature 
see label)
MIN
MAX
AUTO
°C
Using the push/turn control select the 
symbol 
MAX
-Symbol anwählen. 
Hold  down  the  SET  key  and  set  the 
protection  temperature  using  the 
push/turn control.
13.2.2 Undertemperature protection 
MIN
Range: from  10°C  below  minimum 
temperature  of  oven  to  10°C  above 
nominal  temperature  of  oven  (for 
nominal temperature see label). 
The low alarm cannot be programmed 
above the value set as high alarm.
Where  no  undertemperature 
protection is required, this has to be 
set to the lowest temperature.
MIN
MAX
AUTO
°C
Using the push/turn control select the 
symbol 
MIN
.
Hold  down  the  SET  key  and  set  the 
protection  temperature  using  the 
push/turn control.
Note: 
The temperature monitor can be set independently of the operating mode.
During ramp operation the monitor temperature must always be set sufficiently far above the maximum 