Operating instructions
PERFECT page 30
Going outside tolerance margin = ASF alarm
Going outside the selected tolerance margin of the setpoint (in the example 150°C +/-10°C), for example 
through opening the oven door during operation, triggers the alarm.
Triggering the ASF alarm is indicated by flashing 
AUTO
 and   -symbol.
If the sounder is switched on in SETUP, the ASF alarm is additionally signalled by an interrupted tone. By 
pressing the SET key the sounder can be switched off temporarily until the next occurrence of an alarm 
event.
160°C
140°C
ASF alarm
ASF activated
ASF activated
160°C
140°C
ASF-alarm
ASF activated ASF activated
85°C
65°C
Again within tolerance margin = ASF alarm switched off
The automatic temperature monitor alarm is switched off automatically as soon as the selected tolerance 
margin of the setpoint (in the example 150°C +/-100°C) is reached again.
160°C
140°C
ASF alarm
ASF activated
ASF activated
160°C
140°C
ASF-alarm
ASF activated ASF activated
85°C
65°C
ASF set to 
+/-10°C in 
SETUP
ASF set to 
+/-10°C in 
SETUP