Operating instructions
PERFECT page 35
15.2  Bus interface RS485
When so ordered, the oven can be equipped at the factory with an RS485 interface instead of the RS232C 
interface. This permits networking of several ovens (up to 16) with a single PC using a common 2-wire circuit. 
The system is operated using the “Celsius 2007” software. A unique device address has to be assigned to 
the oven in sub-menu SETUP, option ADDRESS. This is the address through which the PC communicates 
with the oven. The default setting is ADDRess 0. Using this address each oven can be addressed by the PC 
and programmed.
For this purpose the PC must be equipped with an RS485 interface or must be fitted with an RS232/RS485 
converter. The cabling has to suit the individual location using a screened cable. The maximum total length 
of the cable is 150 m.
A maximum of 16 devices can be addressed on the RS485 bus. A termination resistance of 220 Ohm has 
to be connected to the last device.
RS 485
9-pin serial
12345
69 8 7
1 2 3 4 5 6 7 8 9
not
used
not
used
A not
used
not
used
not
used
not
used
B not
used
RS485
RS232
Converter