Operating instructions
PERFECT page 44
Programming example steriliser
The steriliser (SFP600) has to sterilise at a temperature of 180°C and a medium quantity of load for one 
hour and 30 minutes. By setting a cooling time the load can only be removed after it has cooled down.
sterilisation time
heating time
180°C
cooling time
°C
t=time
1. Select operating mode “Programme“
After holding down the SET key (approx. 3 sec) the current operating 
mode is flashing. Hold down the SET key and select operating mode 
“Programme“ using the push/turn control.
After the SET key has been released, the controller is in the “Programme” 
operating mode.
PRINT
SETUP
2. Edit programme
Hold down the SET key and  turn the push/turn control to  select 
“EDIT”.
After the SET key has been released, the controller is in the programme 
writing mode.
3. Immediate programme start
Hold down the SET key and turn the push/turn control to select INSTANT 
START.
4. Select duration of first ramp segment
Turn the push/turn control to select the time indication.
Hold  down  the  SET  key  and  set  the  time  00:01  using  the  push/turn 
control.
h
5. Set temperature of first ramp segment
Turn  the  push/turn  control  clockwise  until  the  temperature  display  is 
flashing. 
Hold down the SET key and set the required temperature setpoint of  
180 °C using the push/turn control.
˚C
6. Set fan speed for first ramp segment
Turn the push/turn control clockwise until the fan symbol is flashing.
Hold down the SET key and set the fan speed to 100% using the 
push/turn control.
The  fan  motor  is  running  at 
maximum speed.