Operating instructions
PERFECT page 46
14. Select duration of third ramp segment
Using the push/turn control select the time indication
Hold  down  the  SET  key  and  set  the  time  00:01  using  the  push/turn 
control
h
15. Select temperature of third ramp segment
Turn  the  push/turn  control  clockwise  until  the  temperature  display  is 
flashing.
Hold down the SET key and set the required temperature setpoint of 
30 °C using the push/turn control.
˚C
16. Select fan speed for third ramp segment
Turn the push/turn control clockwise until the fan symbol is flashing.
Hold down the SET key and set the fan speed to 100% using the push/
turn control.
17. Set air valve for third ramp segment
Turn the push/turn control clockwise until  the air valve symbol is 
flashing.
Hold down the SET key and set the air valve to 100% using the push/
turn control.
18. Set closure command for third ramp segment
Turn the push/turn control clockwise until a segment closure command 
(e.g. end ) appears.
Hold down the SET key and set SPWT [T] with the push/turn control.
The segment closure command 
SPW [T] ensures that 
sterilisation is terminated 
only after the unloading 
temperature has been reached.
19. Select duration of fourth ramp segment
Using the push/turn control select the time indication
Hold  down  the  SET  key  and  set  the  time  00:01  using  the  push/turn 
control
h
20. Select temperature of fourth ramp segment
Turn  the  push/turn  control  clockwise  until  the  temperature  display  is 
flashing.
Hold down the SET key and set the required temperature setpoint of 
20 °C using the push/turn control.
˚C
21. Select fan speed for fourth ramp segment
Turn the push/turn control clockwise until the fan symbol is flashing.
Hold down the SET key and set the fan speed to 100% using the push/
turn control.