Operating instructions
PERFECT page 5
 15 cm
 min.
 8 cm
 min.
 8 cm
 min.
 20 cm
 min.
3  Installation facilities (accessories)
The oven can  be  placed  on  the floor or  on  a 
bench  (working  surface).  It  is  important  that 
the oven is set up accurately horizontally; the 
door  may  have  to  be  adjusted  (see  Section 
„Maintenance“)
The spacing from the back of the oven to the 
wall should be at least 15 cm. The spacing to 
the  ceiling  must  not  be  less  than  20  cm  and 
that at the side to the wall not less than 8 cm. 
Generally  it  is  essential  to  have  adequate  air 
ventilation around the oven.
Model 800 is mounted on castors. The front castors pivot and can be locked. In order to ensure the stability 
of the oven the front castors must always be set facing towards the front.
Information on accessories will be found in our leaflet or on our internet page www.memmert.com.
Please note the installation instructions for our accessories.
3.1  Subframe (accessory)
Oven models 500 to 700 can be mounted on a subframe
3.2  Wall bracket (accessory)
Oven models 200 to 700 can be wall-mounted using the wall bracket. The wall 
bracket  is  factory-fitted  with  a  fire-resistant  plate.  The  size  and  length  of  the 
screws used and of the corresponding dowel plugs depend on the total weight 
(oven plus load) and vary with the condition of the wall.
3.3  Stackable version (accessory)
Two  ovens  of  the  same  model  size  can  be  stacked  on  each  other.  Note  that 
the  oven  with  the  lower  working  temperature  must  always  be  placed  at  the 
bottom.
Foot locators  have  to be  fitted  on  the  bottom  oven.  (Model  700  can  only  be 
stacked using an intermediate frame.)
Take off cover of bottom oven•
Place drill jig (supplied with foot locators) into the inverted cover at the  •
  back
Mark holes and drill 4.2 mm dia.•
Screw the foot locators to the top of the cover using the screws and nuts •
  supplied
Re-fit the cover•