Actual Volumetric Flow Rate
1) To obtain actual volumetric flow rate if the inlet
temperature is different from standard
temperature,
read the flow rate (Q) from the curve corresponding
to
the corrected DP.
Calculate
the
viscosity
at the
flowing temperature
using the viscosity equation for the flowing gas.
Then
calculate the viscosity correction factor
using
µcf =


Multiply Q by the viscosity correction factor µcf.
This
product will give the flow rate of a particular
gas at
the
actual flowing
conditions.
3)
At flowing inlet
temperature
equal to the
standard
temperature,
read the
curve/table
directly in
actual
volumetric flow rate.