User Manual

15
Temp
°F
+0
+1
+2
+3
+4
+5
+6
+7
+8
+9
50
60
70
80
90
100
110
120
130
140
150
1.0392
1.0192
1.0000
0.9815
0.9636
0.9464
0.9298
0.9137
0.8982
0.8833
0.8688
1.0372
1.0173
0.9981
0.9797
0.9619
0.9447
0.9282
0.9122
0.8967
0.8818
0.8674
1.0352
1.0153
0.9962
0.9778
0.9601
0.943
0.9265
0.9106
0.8952
0.8803
0.8659
1.0332
1.0134
0.9944
0.976
0.9584
0.9414
0.9249
0.909
0.8937
0.8789
0.8645
1.0311
1.0115
0.9925
0.9742
0.9567
0.9397
0.9233
0.9075
0.8922
0.8774
0.8631
1.0291
1.0095
0.9906
0.9725
0.9549
0.938
0.9217
0.9059
0.8907
0.876
0.8617
1.0271
1.0076
0.9888
0.9707
0.9532
0.9364
0.9201
0.9044
0.8892
0.8745
0.8603
1.0252
1.0057
0.987
0.9689
0.9515
0.9347
0.9185
0.9028
0.8877
0.8731
0.8589
1.0232
1.0038
0.9851
0.9671
0.9498
0.9331
0.9169
0.9013
0.8862
0.8716
0.8575
1.0212
1.0019
0.9833
0.9654
0.9481
0.9314
0.9153
0.8998
0.8847
0.8702
0.8561
Troubleshooting
Problem
Probable Cause Remedy
Low or
High
DP
Indication
Insufficient or improperly sized straight pipe
downstream and/or upstream of LFE.
Use 10 diameters of straight pipe
upstream and 5 diameters of straight
pipe
downstream
of LFE. Pipe size
should be same as LFE outlet size,
e.g. ½" NPT on LFE means
10
diameters of ½ pipe.
One or both differential pressure
connection
taps
plugged.
Clean or check instrument
connecting line.
If pulsation dampener is used, check
stones
(Model 50 MR2 and 50 MC2
only)
.
If plugged, replace with matched
pair
(Meriam part #A-31650).
Leak in line between LFE and readout
device
.
Detect and repair.
Large-volume and/or unequal-volume
connecting lines to readout device.
Use small-volume and
equal
volume
connecting lines to readout
device. See Installation on page
4.
Piping reducers at inlet and/or outlet. Do not use reducers immediately
before or after
LFE.
Pulsating /
Irregular Reading
Irregular flow pattern entering LFE. Use at least 10 diameters of straight
pipe
upstream of
LFE.
Leak in system line. Detect and repair.