Owner manual

18
STEP BY STEP: P5, SETPOINT OPTIONS
Step Action Display
1
Gauge should be in the normal Measure Mode. Normal Pressure Display.
Press the PRGM/Enter key.
2
If the lockout is active, the gauge will now prompt for entry of
the lockout code. Refer to “STEP BY STEP: ENTERING A
LOCKOUT CODE (When Prompted)” on page 22.
The PRGM annunciator shows at the top of the
display, and the display shows the register name
P0”.
Press the Up or Down keys repeatedly to scroll
to the desired register.
3
The Zero function (Up and Down) will reset the scroll to P0
and allow selection to continue.
Display shows “P5”.
4
Press PRGM/Enter to open the register.
The display shows the current value.
Press the Up or Down keys repeatedly to scroll
to the desired register value.
5
The Zero function (Up and Down) will reset the scroll to
default and allow editing to continue.
All available choices are scrolled.
6
Press the PRGM/Enter key to accept the value.
The value is accepted, and the register is closed.
Display shows “P5”.
7
Press the Backspace key to activate the register
setting and return to Measure Mode.
Normal Pressure Display.
Notes
1. After steps 2 and 3, the Backspace key will abort the process and return to Measure Mode.
2. After steps 4 and 5, the Backspace key will abort the edit and close the register.
3. After step 6, the PRGM/Enter key will again open the register for edit.
4. During any programming operation, if there is no keypad activity for approximately 1 minute, the
operation is aborted and the gauge returns to Measure Mode unchanged.