Notes on RTD wiring 
1.  For maximum accuracy use 3-wire or 4-wire configurations with DIN Class A RTDs. 
2.  When using a 2-wire RTD, keep the lead length between the alligator clips of the general- 
 purpose connector and the RTD element as short as possible. 
3.  In 3-wire and 4-wire configurations, the wire size and length should be identical for all leads. 
Option 2: RTD Measure Cable 
IMPORTANT: Alligator clips on the connector/cable assembly must all be connected to something for 
readings to take place! 
RIO Measure Cable 
P/N A900028-90500 general-purpose connector is used for connecting to RTD wires or terminals to obtain 
temperature measurements directly from RTD devices. One pair of white and one pair of red Alligator clips 
are suitable for connecting to 2-, 3-, or 4-wire RTDs. 
 2-wire RTDs 
    Connect the two Red clips to one wire/terminal of the 
    RTD. Connect the two White clips to the other 
  wire/terminal. 
 3-wire RTDs 
    3-wire RTDs have two wires/terminals that are the same 
    color. Connect the two Red clips to the wires/terminals of 
    the same color. Connect the two White clips to 
    the remaining wire/terminal. 
 4-wire RTDs 
    4-wire RTDs have two wires/terminals of one color and two 
    of another color. Connect the Red clips to each of two 
    wires/terminals of the same color. Connect the White clips 
    to each of two wires/terminals that are the other color. 
Note: Restrictions apply for use of Intrinsically Save RIO4000 modules / MFT units in Hazardous 
Areas. RTDs and / or RIO4000 modules must only be removed or replaced in non-hazardous areas. 
RTDs, P/N A900442-00026 connectors and P/N A900028-90500 User Cable must only be connected to 
the RIO4000 in a non-hazardous area. See the Intrinsically Safe Control Document in the Appendix 
for more details. 
2.2  Measure Mode User Interface 
2.2.1  Install the RIO4000 module in one of the MFT bays and turn on the MFT. If no RTD is connected to the 
RIO module, the corresponding Sx line on the display will indicate “No Probe.” Be sure the RTD type is 
correct - if not correct, go to the RTD Sensor Module Setup display to change the type. 