MFT CONFIGURATION SETTINGS 
Press the Settings key to review the preferences on the MFT for Users, Measurements, Applications, Lockouts, 
Clock/Timers and Other options This will display the Settings menu. Note: Not all options will be active on all 
MFT models.  
The Settings menu shows the current settings for the MFT and allows the user to make adjustments through the 
use of the soft keys: 
Users (active only on 21CFR Part 11 capable MFTs) 
Displays the approved Users for Configuration changes and Calibrations – See the MFT 21CFR Part 11 User’s 
Manual for more information. 
Measurements  
Allows user to select the default Damp mode and to select display units 
for Sensor Bays 1, 2, and 3 plus the integral volts / current display. 
Damping 
Allows user to select between no Damping, Smart Damp, or 
exponential Damp. Smart Damp is the factory default mode and 
provides the most stable yet responsive display of measured values. 
This can be changed to exponential Damp to provide the user with 
control of the averaging time constant, or damping can be completely 
removed. Move the cursor to the Damping menu line of the 
Measurements Menu Screen and press the Chng Soft key to scroll 
through the available options and select the desired damping. 
Damping method selected here becomes the normal damping mode. 
Units 
The Units menu line of the Measurements Menu Screen allows the 
user to scroll through and select the available units of any of the 
three Sx sensor module bays and the integral MFT volts/current 
meter. Move the cursor to the desired line and press the Chng Soft 
key to scroll through the available options. 
Applications  
Allows user to select special applications available on the MFT. Follow the on-screen prompts and directions to 
utilize these Applications. 
Digital Poll (addresses 0 – 15) 
Digital Polling is a special feature accessible through the MFT’s “Applications” menu option. Select the 
“Digital Poll” menu option to launch a polling operation that includes address 0 and ends with address 15. 
All addresses on the loop will be displayed on the MFT’s screen. The address of interest can be selected 
from the list to launch HART communication with that device. Digital Poll can only be launched from this 
location in the handheld’s menu structure. 
Settings Main Menu Screen 
Configuration # 
Current Settings 
 Users 
 Measurements 
 Applications 
 Lockouts 
 Clock/Timers 
 Other 
Up Down Chng Back 
Configuration 
Current Settings: 
 Damping:   Smart 
 S1 Units:    inW20C 
 S2 Units:    PSI 
 S3 Units:    deg F 
 S1 Units:    inW20C 
 IV Units:    mA 
Up Down Chng Back 