Vacuum and Absolute Pressure Calibration 
Vacuum calibrations can be performed with an x
DN sensor module by venting the high pressure connection to 
atmosphere and applying the vacuum to the low pressure connection. Atmospheric pressure is the reference pressure 
for all vacuum measurements. 
Absolute pressure calibrations use a x
AI sensor module. These sensor modules have an internal absolute zero 
reference. Simply connect the pressure (or vacuum) source to the single port on the module. 
The following diagram shows the connection for both a vacuum and an absolute pressure measurement. If the 
process maintains a stable pressure, it can be used as the pressure source in some calibrations. Process 
compatibility and pressure ranges of the sensor modules need to be considered for each application. 