will appear and the cursor will be pointed to the offending area of the setup screen. Once the setup is correct, 
the following screen will appear: 
Manual Cal: choose this option to perform free style calibration checks by 1) applying a pressure of interest 
or 2) by selecting a simulation value manually. The user creates this calibration documentation as he goes. 
Save as many free-style points as needed to complete the calibration documentation session. Selecting 
Manual Cal for a pressure calibration will cause one of the following displays to appear. 
  Manual Cal Screen for        Manual Cal Screen for 
    Conventional Devices        HART Devices 
Apply the desired input signal to the MFT and device under test, wait for the value to stabilize and press the 
Save soft key to save the displayed data to memory. The MFT will provide “Saving Data Point…” feedback 
on the display, show “1 Points Saved,” and prepare to accept the next point. 
Apply the next desired input signal and press the Save soft key when satisfied with the input signal. Repeat 
as needed to complete the As Found points. Press the Done soft key to end the As Found portion of the 
documenting session. 
If the MFT needs to provide simulation of an input signal to the device under test, the Meas / Sim line will 
have been pointed to a simulation module (such as the RIO4000, TIO0110 or VMA0055 modules). In these 
cases, starting the Manual Calibration will prompt the user to Enter Output Value using the numeric entry 
methods on the MFT. Once the first point is saved to memory, the MFT will ask if the user wants to simulate 
another point. Selecting Yes continues the Manual Calibration while selecting No ends the session. 
Create (Manual) Procedure: chose this option to develop a structured calibration procedure through the 
MFT user interface. Once completed, the procedure will either 1) prompt the user with a pressure value to 
apply or 2) start a simulation procedure using the simulation module selected. The Create Procedure set 
up screen is shown below. To change values simply move the cursor to the desired menu option and press 
the Select soft key. Use the MFT’s user interface to change or input the values desired. 
Cal 
Manual Cal Proc 
 Manual Cal 
 Create Procedure 
Up Down Select Back 
Cal 
Device: DPT-1003 
 S1: 0.00 inW20C 
 IV: 4.006 mA 
 0 Points Saved 
Save     Done 
Cal 
Device: DPT-1003 
 AO: 4.004 mA 
 Pres: 0.07 inH2O 
 IV: 3.991 mA 
 S1: -0.00 inW20C 
0 Points Saved 
Save     Done 