Local Indicator or Fieldbus Device Calibrations and Documentation 
To document the As Found and As Left condition of local indicating devices or Fieldbus devices, select 
Enter for the Output option on the main Cal Setup screen and then select Begin Calibration. The following 
screen will appear. 
Where: 
S1 (sensor module bay Sx): shows the input value applied to the device as 
 measured by sensor S1. 
Enter (local indicator of fieldbus device): shows output value from the 
 device under test that was keyed in by the user. This value is read 
 from a local indicator or an external meter or relayed from the 
 control room. 
To enter a value in the Enter line, press the Entry soft key to activate the keyed entry 
system. 
Use the numeric input feature on the MFT’s keypad to enter the local indicator 
reading or the output reading from the fieldbus device. 
Or, use Inc and Dec Soft keys enter the first number of the value being input. 
Press the Æ soft key to move to the next place on the entry screen. 
Repeat until complete value is entered and press the Done Soft key. 
The user will be prompted, “Save New Data?  Yes No”. Select Yes if correct. 
When satisfied with the value, press the Save key to save this calibration point 
and proceed to the next point. 
Continue until the As Found calibration procedure is complete 
Press the Done key to complete the “As Found” portion of the calibration and the 
 As Found Complete Screen will appear (see below). 
 Cal 
 Manual Calibration 
 S1 :10.000 inW20C 
 Enter: 0.000 inW20C 
 0 Points Saved 
 Save  Entry Done 
 Cal 
 Enter Xmitter Value 
 9.910
 Å Æ  Inc Done 
 Cal 
 Manual Calibration 
 S1 :10.000 inW20C 
 Enter: 9.910 inW20C 
 0 Points Saved 
 Save  Entry Done 