Manual
56
FUNCTION and then the PgDn soft key. To retreat to the prior page, press FUNCTION and then PgUp. 
Once engaged, the PgUp and PgDn soft keys remain active until the Up or Down soft keys or FUNCTION 
is pressed again. 
List / Edit Configs: 
Lists all stored HART configurations by Tag Number. Once a 
desired tag is located with the cursor, pressing the Select soft key 
displays header information to help the user confirm the identity 
and origin of the configuration. Pressing Select again opens the 
configuration for review and editing as needed. Any configuration 
in the list can be sent to another HART device of the same 
manufacture and model number. This feature enables fast 
commissioning of replacement devices, cloning of existing 
systems, or re-configuring for changes in process conditions or 
batch runs. New entries to this list are loaded to the bottom of the 
list. If multiple entries for one device are shown, the most recently 
saved configuration is always listed last. 
Entries in this list may have been saved directly from a connected HART device, created in the MFT using 
the Create Configs function, copied from another configuration and renamed, or created by editing an 
existing file. MFT will also list configurations downloaded from the separately available Meriam Device 
Management System software (see www.meriam.com for more information). 
Create Configs: 
The Create Configs function allows the user to configure a HART device file in Offline Mode for sending to 
the intended device later when connected in the Online Mode. Use the soft key controls to move through 
the list and select the device model number required. The menus prompt the user to make the necessary 
configuration selections, tag the file for later retrieval and use, and edit configuration lines as required. This 
function can be locked out. See the Lockout Section of this manual for more information. 
Delete Individual Configs / Clearing Configuration Memory: 
Individual stored Configurations may be deleted with MFT Del soft 
key functions (see example at right). A delete confirmation screen is 
provided to prevent unintended deletions. The Del key only hides the 
record from the MFT screen; it does not clear memory space. 
To clear all
 configurations from memory (including active and hidden 
configs), enter the HART Offline Menu Screen and select the 
Delete All Configs option. The MFT will ask “Delete ALL Configs… 
ARE YOU SURE???” Select Yes to clear all
 stored configurations 
from memory. 
HART ONLINE MODE 
Once a HART device is located, the MFT enters the Hart Device Online Screen as shown below. The word 
“Hart” appears in the top left of LCD. The   symbol will “flash” when HART communications have been 
established. 
Note: The screen at right illustrates a Rosemount 1151 rev. 
5 HART display. Screen configuration may vary depending 
on the HART device connected. For more information, 
please refer to the owner’s manual for the HART device 
under test. 
Hart   
1151: Pump 8 
Device Config 
 Pres:  0.000 InH20 
 AO:  4.000 mA 
 LRV:  0.000 InH20 
 URV:  104.8 InH20 
 % of Range: 0.000% 
 AO Fixed 
U
p
 Down Select Back
Offline List / Edit 
Confi
g
s Screen
Offline    
List/Edit Configs 
PT-105B 
DPT-201C 
TT-312 A 
 TT-312 A 
 FE-201 
 FE-201 
 PT-6174 
Up Down Select Back 
Delete Individual Configuration 
Offline    
List/Edit Configs 
 PT-105A 
 Device ID 6121486 
 Single Var 
 Endress + Hauser 
 Cerabar S  
 Complete Config 
 Uses DOF 11070702 
 81 Variables 
 02/21/03 4:50 PM 
Copy Del Select  Back 