ROV-II operation manual
14 Document 354-MN01-ENG Rev. C
Figure 6
Cell to Ag/AgCI master
ROV-II calibration wiring schematic
Calibration
voltmeter display
at 200 mV DC setting
Seawater solution
Non-metallic bucket
Step +VE —VE DC scale Expected result
01 Master Reference electrode 1 / Pin 2 200 mV 0 mV ± 5.0 mV
02 Master Reference electrode 2 / Pin 3 200 mV ±5.0 mV
03 Master Tip / Pin 1 2 V -3 V ± 0.1 V
Ag/AgCI master
electrode
Note:
This calibration
procedure is
only performed
by qualied
Polatrak
technicians
before
shipment or
during repair.