User Manual

Model 2200P Page 30
9.3.5 After the "ESCAPE" procedure it is necessary to do the following:
a) Calibrate the temperature as described in Section 7.4.
b) Tune the analog outputs and the temperature output span to suit your particular
application and loop. See Sections 7.5, 7.6 and 7.7.
c) Leave the Utility Menu by pressing RUN.
d) Calibrate the pH, as described in Section 5.2.
e) Set up the user values for:
• Output High and Low – See Section 5.4
• Control Relay A – See Sections 5.5 and 5.6
• Control Relay B – See Sections 5.7 and 5.8
• Alarm Relay – See Sections 5.10 and 5.11
• The cycling parameters if so equipped – See Section 5.14
f) After all the above operations are performed the STATUS will turn green and
status 0 should be obtained.
g) Bring the unit on-line for measurement and control by pressing RUN.
h) The unit may be tested now, using the TEST menu item (see Section 5.12), or
the built-in self-testing feature (see Section 9.2.5).
9.4 Printed Circuit Board Replacement
9.4.1 a) To replace printed circuit boards or relays shut off all power to the 2200P,
including any independent power to the relay contacts. Make a record of the
external wiring, then disconnect the wires. Unplug the ribbon cable connector.
b) The power supply circuit board is fastened to the back of the enclosure by four
screws, remove the screws and the board is free. Reverse the procedure to
mount a replacement board.
c) The microprocessor circuit board is located on the swing-out assembly behind
the door. Swing the assembly out, lift it up to unseat the lower hinge pin (the
upper hinge pin is spring loaded.) The assembly is now free. The circuit board
is fastened to the front panel by five screws. Remove the screws to release the
circuit board.
9.5 Customer Service
9.6.1 If a problem has not been resolved with the above procedures, a telephone
consultation with your AquaMetrix representative, or directly with AquaMetrix will
provide the answer.
AquaMetrix Inc. Tel: (800) 742-1413
22-121 Granton Drive (905) 763-8432
Richmond Hill, Ontario Fax: (905) 763-9480
L4B 3N4 Canada Email: support@aquametrix.com