Operation Manual

OmegaGTPlus / BetaGTPlus Instrument description
12
Leakage LN-PE or EB-
PE too high!
Are you sure to
proceed? (Y/N)
Dangerous leakage current (higher than 20 mA) will flow if
power would be connected to the tested appliance.
Select YES or NO with Y or N key.
Proceed with testing only if all safety measures have
been taken. It is recommended to perform a thorough
earth bond test on the PE of the appliance before
proceeding with the test.
Measurement aborted!
Differential leakage too
high!
An exceptionally high leakage current (higher than about 5
mA) was measured through PE test terminal or PE socket
during the pre-test or test.
Leakage test probe-PE
too high!
An exceptionally high leakage current (higher than about 5
mA) was measured through S test terminal during the pre-
test.
External voltage on test
socket too high!
DANGER!
Voltage on test socket or LN/PE terminals is higher than
20 V (AC or DC) approximately!
Disconnect the appliance under test from the instrument
immediately and determine why external voltage was
detected!
External voltage on test
probe too high!
DANGER!
Voltage on test probe (S) is higher than 25 V (AC or DC)
approximately!
Disconnect the test probe from the appliance and
determine why external voltage was detected!