Operation Manual
OmegaGTPlus / BetaGTPlus Main menu and test modes
22
Note:
Ì Use the test probe for class II
appliances and class II parts of class I
appliances!
Substitute leakage current measurement - Class II
Notes:
Ì If any of the earth bond resistance,
insulation resistance, or substitute
leakage current measurements have
failed in selected autotest sequence, the
differential current measurement is
prohibited!
Ì If leakage current measurement fails,
test sequence will be finished.
Ì The instrument automatically changes
L/N polarity of connected appliance.
Differential current measurement
Note:
Ì The instrument has possibility to
measure substitute and differential
leakage and thus not need to measure
PE leakage (see * on organizer
flowchart).
Leakage current
Note:
Ì If any of the earth bond resistance,
insulation resistance, substitute leakage
current or leakage current
measurements have failed or were
skipped in selected autotest sequence,
the touch leakage current measurement
is prohibited!
Ì The earth bond test probe shall be used