Operation Manual

OmegaGTPlus / BetaGTPlus Main menu and test modes
30
Keys:
À / ¿
F1 (PgUp)
F2 (PgDown)
Select the appliance that should be retested.
START
Starts execution of the autosequence for selected appliance (retest).
See 3.1.3 Performing autotest sequences.
ENTER
Opens more detailed information of appliance results and data.
ESC
Returns to Main menu.
Appliance tickers
Each appliance is marked with a ticker at the right end of the line. The tickers help to
faster find the appliance to be retested from the offered list.
Their meaning is following:
PC
The stored appliance data was uploaded from PC
PC9
The stored appliance data was uploaded from PC
It was already used for a new Project autotest measurement.
I
The stored appliance data was performed with the instrument and stored.
I9
The stored appliance data was performed with the instrument and stored.
It was already used for a new Project autotest measurement.
Keys in appliance view:
À / ¿
F1 (PgUp)
F2 (PgDown)
Scroll displayed items.
ENTER
Opens more detailed information of appliance results and data.
ESC
Returns to Main menu.
F5 (MORE)
Switches between different appliance data views.
View results menu examples
Notes on retesting the selected appliance:
Ì The Autotest sequence will be the same as it was for the selected Project
autotest.
Ì Exactly same appliance data (except time and user) will be offered as stored in
the selected Project autotest.
Ì The new autotest will get an »I« ticker. The original autotest will get a »I9« ticker
(if it was made with the instrument, or PC9 ticker (if it was uploaded from the
PC).