Operation Manual

OmegaGTPlus / BetaGTPlus Appendix A
95
METREL GmbH VDE tester test type card (cont'd)
Code Autotest sequence name and descriptions Limits Barcode
09
702_Kl
_2_IsoI
bs
Testing according to VDE 702.
Class 2 appliance with isolated
accessible conductive parts.
Prüfungen für Isolation und
Berührungsstrom werden eingestellt.
Insulation - S: 2.00 MΩ
Touch leakage: 0.50 mA
09
10
702_Kl
_2
Testing according to VDE 702.
Class 2 appliance without any isolated
accessible conductive parts.
10
11
701_Kl
_1_Iso
Testing according to VDE 701.
Class 1 appliance.
Insulation resistance and substitute
leakage current measurements are
applicable.
Earth bond: 0.30 Ω
Insulation: 1.00 MΩ
Sub leakage: 3.50 mA
11
12
701_Kl
1_Iso_
BLT
Prüfung nach VDE 701.
Class 1 appliance with isolated
accessible conductive parts.
Insulation resistance and substitute
leakage current measurements are
applicable.
Earth bond: 0.30 Ω
Insulation: 1.00 MΩ
Insulation - S: 2.00 MΩ
Sub leakage: 3.50 mA
Sub leakage - S: 0.50 mA
12
13
701_Kl
_1_Ia
Testing according to VDE 701.
Class 1 appliance.
Prüfung für Differenzstrom wird
eingestellt.
Earth bond: 0.30 Ω
Leakage: 3.50 mA
13
14
701_Kl
_1_Ia_
BLT
Testing according to VDE 701.
Class 1 appliance with isolated
accessible conductive parts.
Prüfungen für Differenz- und
Berührungsstrom sind eingestellt.
Earth bond: 0.30 Ω
Leakage: 3.50 mA
Touch leakage: 0.50 mA
14
15
701_Kl
_2_Iso
Testing according to VDE 701.
Class 2 appliance with isolated
accessible conductive parts.
Insulation resistance and substitute
leakage current measurements are
applicable.
Insulation - S: 2.00 MΩ
Sub leakage - S: 0.50 mA
15
16
701_Kl
_2_Ibs
Testing according to VDE 701.
Class 1 appliance.
Prüfung für Berührungsstrom wird
eingestellt.
Touch leakage: 0.50 mA
16