Operation Manual

MI 3309 BT DeltaGT Technical specifications
19
3.5 Polarity test
Powered by ............................... battery or mains
Test voltage .............................. <50 V AC
Detects ...................................... PASS, L OPEN, N OPEN, PE OPEN, L-N CROSS, L-N
SHORT, L-PE SHORT, N-PE SHORT, MULTIPLE FAULT.
Test terminals ........................... Test socket IEC test terminal
3.6 Differential leakage current
Range Resolution Accuracy
0.00 mA 19.99 mA
0.01 mA
(5 % of reading + 5 digits)
Apparent power Indication only
Powered by ....................... mains
Pass levels: ....................... 0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.00 mA, 2.25
mA, 2.50 mA, 3.50 mA, 4.00 mA, 4.50 mA, 5.00 mA, 5.50 mA,
6.00 mA, 7.00 mA, 8.00 mA, 9.00 mA, 10.00 mA, 15.00 mA, ----
mA
Test duration: .................... 2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, --- s
Frequency response: ........ complies with EN61010-1 Figure A1
Test terminals ................... Test socket
Additional error ................. 0,01 mA/A
* Measurement is limited to 120 s if (I
load
> 10 A).
3.7 Touch leakage current
Range Resolution Accuracy
0.00 mA 7.00 mA
0.01 mA
(10 % of reading + 5 digits)
Apparent power Indication only
Powered by ..................... mains
Pass levels: ..................... 0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.00 mA 2.25
mA, 2.50 mA, 3.50 mA, --- mA
Test duration: .................. 2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, --- s
Frequency response ....... complies with EN61010-1 Figure A1
Test terminals ................. Test socket – EB/S test probe
* Measurement is limited to 120 s if (I
load
> 10 A).