Operation Manual
MI 3309 BT DeltaGT Technical specifications
21
3.9 Power
Apparent power
Range Resolution Accuracy
0.00 kVA 4.00 kVA
0.01 kVA
(5 % of reading + 3 digits)
Current
Range Resolution Accuracy
0.00 A 16.00 A
0.01 A
(5 % of reading + 3 digits)
Powered by ..................... mains
Test duration: ............................ 2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, --- s
Test terminals ........................... MAINS terminal
* Measurement is limited to 120 s if (I
load
> 10 A).
3.10 TRMS Voltage
Range Resolution Accuracy
80 300 V
1 V
(2 % of reading + 2 digits)
Result type ................................ True r.m.s.
Nominal frequency range .......... 0 Hz, 50 Hz 60 Hz
Frequency accuracy ................. Indication only
Test terminals ........................... MAINS terminal
3.11 Clamp current
True RMS current using 1000:1 current clamp (A 1283)
Range Resolution Accuracy*
0.00 mA 9.99 mA
0.01 mA
(5 % of reading + 10 digits)
10.0 mA 99.9 mA
0.1 mA
(5 % of reading + 5 digits)
100 mA 999 mA
1 mA
(5 % of reading + 5 digits)
1.00 A 9.99 A
0.01 A
(5 % of reading + 5 digits)
10.0 A 16.0 A
0.1 A
(5 % of reading + 5 digits)
*It does not consider accuracy of current transformer.
Pass levels: ....................... 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.00 mA, 2.25 mA, 2.50
mA, 3.50 mA, 4.00 mA, 4.50 mA, 5.00 mA, 5.50 mA, 6.00 mA,
7.00 mA, 8.00 mA, 9.00 mA, 10.00 mA, 15.00 mA, ---- mA
Test duration [s]: ....................... 2, 3, 5, 10, 30, 60, 120, none
Test terminals ........................... Clamp inputs ↔ PE (terminal), LN (terminal)
Temperature coefficient outside reference temperature limits is 1 % of measured value per
C