Operation Manual
MI 3309 BT DeltaGT Single test
45
5.2.9 Touch leakage
This test determines the current that would flow if a person touches accessible conductive
parts of the DUT.
The instrument measures:
The touch leakage current flowing through the S/EB probe into earth.
The DUT can be powered from the test socket or directly from the installation (fixed
installed equipment).
Figure 5.30: Touch leakage menu
Test parameters for touch leakage current measurement
OUTPUT Test voltage [MAINS voltage]
LIMIT
Maximum current [0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.00 mA,
2.25 mA, 2.50 mA, 3.50 mA, --- mA]
TIME Measuring time [2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, --- s (continuous
measurement)
Test circuits for touch leakage current measurement
Figure 5.31: Measurement of touch leakage current