Operation Manual
MI 3309 BT DeltaGT Single test
46
Figure 5.32: Measurement of touch leakage current on a fixed installed DUT
Touch leakage current measurement procedure
Select the TOUCH LEAKAGE function.
Set the test parameters.
Connect device under test to the instrument (see Figure 5.31 and Figure 5.32).
Press the TEST key for measurement. To stop continuous measurement press
TEST key once again.
Store the result by pressing MEM key (optional).
Figure 5.33: Examples of touch leakage current measurement results
Displayed results:
Main result I ........... Touch leakage current
Sub-result P............Apparent power
Notes:
For this test the instrument must be connected to the mains voltage.
During the test, mains voltage is connected to the DUT. If DUT contains moving
parts, make sure that it is safely mounted or protected to prevent possible danger to
the operator or damage to the DUT or surrounding environment!
Consider any displayed warning before starting measurement!
The instrument automatically changes L and N polarity of connected DUT during
the test.
Measurement can be aborted by pressing ESC key.