Operation Manual
MI 3309 BT DeltaGT Autotest sequences
58
Note:
The leakage current limit is automatically set on base of set appliance power.
There are no special options to be set if other Appliance type is set.
Figure 6.9: VDE Organizer – Substitute leakage result screen
There are no special options to be set in the Substitute leakage result screen.
6.1.1.6 Differential leakage measurement
The measurement is described in chapter 5.2.8 Differential leakage.
Figure 6.10: VDE Organizer – Differential leakage starting screen
Options if Appliances with heating elements is set:
/
Sets power of heating elements.
Note:
The leakage current limit is automatically set on base of set appliance power.
There are no special options to be set if other Appliance type is set.
Figure 6.11: VDE Organizer – Differential leakage result screen
There are no special options to be set in the Differential leakage result screen.
6.1.1.7 Insulation resistance - P measurement
This test is offered if it is applicable according to the VDE Organizer setting.
The measurement is described in chapter 5.2.4 Insulation resistance - P.
Figure 6.12: VDE Organizer autotest – Insulation resistance - P starting screen
Options in Insulation resistance-P starting screen:
NO
To be set if insulation test is not applicable
YES
To be set if insulation test is applicable
500 V
Standard test voltage