Operation Manual

MI 3309 BT DeltaGT Autotest sequences
59
250 V
To be set if overvoltage protection devices are installed or
SELV/ PELV protection measures.
Figure 6.13: VDE Organizer – Insulation resistance - P result screen
Options in Insulation resistance - P result screen:
NEXT
Proceeds with the next measurement.
REPEAT
Repeats the test (use in case of multiple isolated / SELV / PELV
accessible points). Lowest result will be stored.
6.1.1.8 Selection of touch leakage current test method
The touch leakage test is offered if it is applicable according to the VDE Organizer setting.
If more test methods are possible the applicable touch leakage current test method must
be set first.
Options are Touch leakage and Substitute leakage-P.
Figure 6.14: Touch leakage current test method selection screen
6.1.1.9 Substitute leakage - P current measurement
The measurement is described in chapter 5.2.6 Substitute leakage - P.
Figure 6.15: VDE Organizer – Substitute leakage - P starting screen
There are no special options in the Substitute leakage - P starting screen.
Figure 6.16: VDE Organizer – Substitute leakage - P result screen
Options in Substitute leakage - P result screen:
NEXT
Proceeds with the next measurement.
REPEAT
Repeats the test (use in case of multiple isolated / SELV / PELV
accessible points). Highest result will be stored.