Operation Manual
MI 3309 BT DeltaGT Appendix B – Pre-programmed autotests (GER)
85
Pre-programmed autotest sequences table (cont’d)
Autotest shortcut code A05 A06 A07 A08
Kl_2_Iso Kl_2_Ibs Kl_1_IsoIa
Kl_1_IsoIaBLT
Visual test
Earth
continuity
Output
200 mA 200 mA
Limit
0.30 0.30
Time
5 s 5 s
Insulation
Output
500 V 500 V
Limit
1.00 M 1.00 M
Time
5 s 5 s
Insulation
(probe)
Output 500 V
500 V
Limit
2.00 M
2.00 M
Time 5 s
5 s
Sub. leakage
Output
Limit
Time
Sub. leakage
(probe)
Output 40 V
Limit 0.50 mA
Time 5 s
Differential
leakage
Output
230 V 230 V
Limit
3.50 mA 3.50 mA
Time
180 s 180 s
Touch
leakage
Output
230 V
230 V
Limit
0.50 mA
0.50 mA
Time
180 s
180 s
Functional test
Power*
Output 230 V 230 V 230 V 230 V
Limit
Time 180 s 180 s 180 s 180 s
TRMS clamp
current
Output
Limit
Time
Polarity test
* optional