5 Single test
In a Single test mode individual tests can be performed. This is especially helpful for
troubleshooting.
5.1 Performing measurements in single test mode
Select appropriate test in Single test menu.
Figure 5.1: Single test menu
Keys:
Selects a single test.
/
Enters Single test measuring menu.
TEST
Returns to Main menu.
ESC
A single test can be started from any Single test measuring menu. Before carrying out a
test the parameters / limits can be edited.
Figure 5.2: Example of single test measuring menu
Keys:
Selects a parameter.
TAB
Changes a parameter / limit.
/
Starts a single test.
TEST
Returns to Single test menu.
ESC
Note:
Last set parameters will be stored automatically.
Single measurements are stored the same way as the autotest results. See chapter 7.1
Saving autotest results for more information.
5.2 Measurements and inspections
5.2.1 Visual inspection
A thorough visual check must be carried out before each electrical safety test.
The following items should be checked:
Inspection of the device under test for sign of damage.
Inspection of the flexible power supply cable for damage.