Operation Manual
MI 3309 DeltaPAT Single test
35
Figure 5.9: Measurement of insulation resistance
Figure 5.10: Measurement of insulation resistance of fixed installed DUTs of Class I
Insulation resistance measurement procedure
Select the INSULATION function.
Set the test parameters.
Connect device under test to the instrument (see figure 5.9 and figure 5.10).
Press the TEST key for measurement. To stop continuous measurement press
TEST key once again.
Store the result by pressing MEM key (optional).
Figure 5.11: Examples of Insulation resistance measurement results
Displayed results:
Main result ............. Insulation resistance