MI 3309 DeltaPAT Single test
Figure 5.16: Substitute leakage menu
Test parameters for Substitute leakage current measurement
OUTPUT Test voltage [30 V]
LIMIT
Maximum current [0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA,
2.00 mA, 2.25 mA, 2.50 mA, 3.50 mA, 4.00 mA, 4.50 mA, 5.00 mA,
5.50 mA, 6.00 mA, 7.00 mA, 8.00 mA, 9.00 mA, 10.0 mA, 15.0 mA, --- mA]
TIME Measuring time [2s, 3s, 5 s, 10 s, 30 s, 60 s, 120 s, --- s (continuous
measurement)]
Test circuits for Substitute leakage current measurement
Figure 5.17: Measurement of Substitute leakage current
Figure 5.18: Measurement of substitute leakage current of fixed installed DUTs