Operation Manual
MI 3309 DeltaPAT Single test
47
If both starting polarities are selected:
Reactivate tested PRCD.
Store the result by pressing MEM key (optional).
or
Select the PRCD test function.
Select test mode.
Set test parameters.
Connect tested PRCD between test socket and IEC connector of the DeltaPAT
(see figure 5.36c), and connect instrument’s MAINS terminal to an external voltage
socket.
Depending on the type of PRCD, it may be necessary to manually switch it on.
Press the TEST key to perform measurement.
If both starting polarities are selected:
Reactivate tested PRCD.
Store the result by pressing MEM key (optional).
Measuring RCD
Select the RCD test function.
Select test mode.
Set test parameters.
Connect the DeltaPAT MAINS terminal to mains socket protected by tested RCD
(see figure 5.36a).
Depending on the type of PRCD, it may be necessary to manually switch it on.
Press the TEST key to perform measurement.
If both starting polarities are selected:
Reactivate tested RCD.
Store the result by pressing MEM key (optional).
Figure 5.37: Examples of (P)RCD single test results
Displayed results:
Main result(s) ......... trip out time(s) at selected starting polarity
U ............................ voltage U
L-PE
5.2.10.2 Automatic (P)RCD test
(P)RCD autotest function is intended to perform a complete (P)RCD analysis (trip-out
times at different residual currents and starting polarity phases) .
(P)RCD autotest procedure
(P)RCD Autotest steps Notes
Select the RCD (PRCD) test function.
Set AUTO mode.
Select test parameters.
PRCD: Connect tested PRCD / device to an external