Operation Manual
MI 3309 DeltaPAT Single test
50
5.2.12 Voltage TRMS
In this function the voltage across the MAINS terminal is measured continuously.
Test circuit for voltage measurement
Figure 5.42: Voltage measurement with the IEC cord
Voltage TRMS procedure
Select the VOLTAGE TRMS function.
Connect the IEC cord to the instrument’s MAINS terminal and to the external
mains socket as shown on figure 5.42.
Store the result by pressing MEM key (optional).
Figure 5.43: Voltage TRMS test result
Displayed results:
Main result ............. Voltage
f.............................. Frequency
Warning:
Only for voltage range from 80 V to 300 V!
5.2.13 Functional test
Functional test is the simplest way to ensure that the appliance is working properly.
Note:
This test should only be performed once the DUT has passed all other tests
applicable to this type of appliances.