Operation Manual

CS 2099 Eurocheck Eurocheck operation
21
Figure 4.5: Performing voltage and frequency calibration using universal test cable
4.13 Continuity and insulation of measurement leads
verification
Continuity of measurement leads
Continuity of universal test cable measurement leads could be verified by voltage
calibration between all 3 installation wires (between L and N, L and PE, N and PE).
Refer to chapter 4.12 for more information. If the tested instrument measures voltage of
around 0 V, at least one of connected leads is broken. By connecting different
measurement lead pairs it is possible to determine which one is broken.
Complete verification can be performed with universal test cable only. Other types of
measuring cables (plug cable, plug commander and tip commander) can be verified
indirectly. First verify universal test cable using procedure for voltage and frequency
calibration (see chapter 4.12). Measure voltages between all measurement lead pairs. If
all results pass, perform any of calibrations using different Eurocheck sockets and
universal test cable. At last perform these calibrations with the other type of cable and
compare the results. If they’re the same or close to each other, the other cable is
acceptable as well.
Notes:
q Suggested calibrations for verifying continuity of plug cable and plug commander
are fault loop (verification of L and PE) and line impedance (verification of L and
N).
q Suggested calibration for verifying continuity of tip commander is low resistance
measurement.
q For quick continuity verification of L and N wires of universal test cable perform
low resistance calibration with any reference value (see chapter 4.11).
P
E
N
L