3.5. Harmonics measurement
The instrument computes harmonics on signals sampled with an A/D converter.
Table 1: Limits of error and resolution in Harmonics measurement:
Range Limits of error Resolution
I
r
, U
r
THD HD on LCD and PC
2 … 100% 0.2 % x U
r
/U (I
r
/I) 0.2 % x U
r
/U (I
r
/I) 0.1 %
Note: THD Total Harmonic Distortion
HD Harmonic Distortion
U
r
U
range
I
r
I
range
3.6. Digital hardware specifications
A/D conversion 14 bit with 128 samples per channel per period (43 - 68 Hz).
3.7. General specifications
Working temperature range
-10 °C … +45 °C
Storage temperature range
Max. humidity
Pollution degree
Protection classification
-20 … 70 °C
85 % RH (0¸40 °C)
2
double insulation
Overvoltage category Voltage inputs: CAT III 600 V, 300 V to gnd
AC power supply CAT III 300 V
Protection degree IP 64
Dimensions 265 x 110 x 18.5 mm
Weight (without accessories) 2 kg
3.8. Maintenance
3.8.1. Batteries
Instrument contains rechargeable NiCd or NiMh batteries. Do NOT replace
with alkaline cells. These batteries should only be replaced with the same type
as defined on the battery cover label or in this manual.
Hazardous voltages exist inside this Instrument. Disconnect all test leads,
remove the power supply cable and switch off instrument before removing
battery compartment cover.
If it is necessary to replace batteries, all four MUST be replaced. Ensure batteries
are installed with the correct polarity; incorrect polarity can damage the batteries
and/or the Instrument.