Operation Manual

MI 2092 Power Harmonics Analyser Section III Operation manual
21
3.1. Password
All programming functions and recorder settings (including the starting & stopping data
logging) are password protected. Unless the password is entered, the various settable
parameters & functions can only be viewed. In all configuration sub-menus, pressing
any edit key (UP, DOWN, LEFT, RIGHT, SELECT, ENTER) will activate password
input procedure. The password must be entered before the SYSTEM sub-menu is
selected.
PASSWORD:
* * * *
Default password
LEFT, SELECT, RIGHT, ENTER
The password is automatically cleared 5 minutes after the last key operation.
3.2. System sub-menu:
This sub-menu allows setting of the password, the serial port baud rate and the
instrument date & time. The fourth choice is to totally reinitialise the Instrument.
ENABLE PASSW.
If the password is enabled ®
CHANGE PASSW.
SER. PORT RATE Or by pressing SELECT GSM / SMS PARAM.
DATE/TIME
LANGUAGE
SYSTEM REINIT. Or by pressing SELECT UPGRADE
CLR.REC.MEM.
Use UP or DOWN keys to select the required menu item, then press the ENTER key.
CHANGE PASSW Enter a new four key combination and repeat it for confirmation.
(The LCD key is not a valid password key)
SER PORT RATE Set the baud rate for serial communication port by using SELECT
key. (from 2400 to 57,600 baud)
GSM/SMS PARAM. See Section VII for modem configuration
Note:
Applicable only if modem option is enabled
DATE/TIME Use the LEFT or RIGHT key to select between Date and Time
and the UP & DOWN keys to set a new date or time.
Only valid date/time values will be accepted.
Press ENTER to confirm the settings or ESC to cancel any changes.
SYSTEM REINIT Clears all settings and sets defaults values as below.
· Recorder START/STOP : MANUAL
· Statistic: ON
· Periodic: ON
· Anomalies: ON, fixed
· Main IP: 1 min
· Power sub IP: 1
· Nominal voltage: 230 V