fig. 13: Recorder (Data Logging) Menu
rec.stat. Present recorder status:
WAIT
Recorder (in AUTO mode) is waiting for start date & time
RUN
Recorder is running
STOP
Recorder (in AUTO mode) has been stopped manually.
Recording aborted.
COMPLETE
Recording completed
buf.stat. Present recorder memory status:
EMPTY
No data in memory
READY
Data present; awaiting download
SAVED
Data present; previously downloaded
Buffer operating status/mode:
lin. Memory in Linear mode
circ. Memory in Roll-over mode
cir./laps Memory in Roll-over mode, current Lap number
start: If the instrument is in Rec.Wait mode and the memory is empty, the
programmed START date & time is displayed.
If instrument is in Rec.Run mode, the actual recording start date & time (as
opposed to programmed) is displayed.
stop: If the instrument is in Rec.Wait or Rec.Run mode, the programmed STOP
date & time is displayed.
If the instrument is in Rec.Stop or Rec.Complete mode, the actual recording
stop date & time (as opposed to programmed) is displayed.
Under certain circumstances, the instrument also displays the reason for
stopping the recording:
MANUAL BREAK
Manual stop in AUTO stop mode
END OF MEM.
Memory full (in linear memory mode)
statist. Statistical Analysis enabled (ON) or disabled (OFF).
anomal. The number of detected and saved Voltage Anomalies.
If currently in a Voltage Anomaly, a blinking arrow points to the number.