BetaGTPlus MI 3304 OmegaGTPlus MI 3305 User Manual Ver. 1.1, Code no.
Distributor: Manufacturer: Metrel d.d. Ljubljanska cesta 77 SI-1354 Horjul E-mail: metrel@metrel.si http://www.metrel.si © 2008 Metrel Mark on your equipment certifies that this equipment meets the requirements of the EU (European Union) concerning safety and electromagnetic compatibility regulations The trade names Metrel, Smartec, Eurotest, Autosequence are trademarks registered or pending in Europe and other countries.
OmegaGTPlus / BetaGTPlus Table of contents Table of contents 1 General description................................................................................................5 1.1 Warnings ....................................................................................................6 1.2 Warning markings on connector panel .......................................................6 1.3 Standards applied ......................................................................................
OmegaGTPlus / BetaGTPlus 5.2.1 5.2.2 5.3 5.3.1 5.3.2 5.4 5.5 5.6 5.7 5.8 5.8.1 5.8.2 5.9 5.10 Table of contents Insulation resistance on class I appliances ..............................................53 Insulation resistance on class II appliances (Insulation S) .......................54 Substitute leakage current........................................................................56 Substitute leakage current on class I appliances .....................................
OmegaGTPlus / BetaGTPlus General description 1 General description Congratulations on your purchase of the instrument and its accessories from METREL. The instrument was designed on basis of rich experience, acquired through many years of dealing with electrictrical appliance test equipment. The multifunctional portable test instrument OmegaGTPlus / BetaGTPlus is intended to perform all measurements for testing the electrical safety of portable electrical equipment.
OmegaGTPlus / BetaGTPlus General description 1.
OmegaGTPlus / BetaGTPlus General description Functionality VDE404-1 VDE404-2 VDE 0701 VDE 0702 Prüf- und Messeinrichtungen zum Prüfen der elektrischen Sicherheit von elektrischen Geräten Allgemeine Anforderungen Prüf- und Messeinrichtungen zum Prüfen der elektrischen Sicherheit von elektrischen Geräten Prüfeinrichtungen für Prüfungen nach Instandsetzung, Änderung oder für Wiederholungsprüfungen Instandsetzung, Anderung und Prüfung elektrischer Geräte Wiederholungsprüfungen an elektrischen Geräten 7
OmegaGTPlus / BetaGTPlus Instrument description 2 Instrument description 2.1 Front panel Front panel Legend: 1..............Mains switch with indication lamp. 2..............Two T16 A / 250 V fuses for instrument protection. 3..............Mains supply cord. 4..............240 × 128 dots graphic matrix display with backlight. 5..............LN and PE sockets for testing the insulation resistance and substitute leakage current of fixed installed appliances.
OmegaGTPlus / BetaGTPlus Instrument description Warning! Do not connect any voltage source on this input. It is intended only for connection of current clamp with current output. Maximum input current is 30 mA! 17............Function keys. 18............USB connector. 19............PC / printer connector. 20............Barcode reader connector. 21............Label printer connector on back side (support for PrintekMobile MtP300 serial printer).
OmegaGTPlus / BetaGTPlus Instrument description Safety pre-tests Before performing a measurement, the instrument performs a series of pre-tests to ensure safety and to prevent any damage. These safety pre-tests check for: • • • • • any external voltage against earth on test socket, excessively high leakage current, excessively high touch leakage current, Short circuit or too low resistance between L and N on appliance, Correctly applied mains voltage to the test socket.
OmegaGTPlus / BetaGTPlus Resistance L – N too high (>30 kΩ). Check fuse and switch. Are you sure to proceed? (Y/N) Resistance L – N low. Are you sure to proceed? (Y/N) Resistance L – N too low. Are you sure to proceed? (Y/N) Leakage LN-PE high. Are you sure to proceed? (Y/N) Leakage LN-PE too high. Are you sure to proceed? (Y/N) Instrument description An excessively high resistance was measured in the fuse pre-test.
OmegaGTPlus / BetaGTPlus Leakage LN-PE or EBPE too high! Are you sure to proceed? (Y/N) Instrument description Dangerous leakage current (higher than 20 mA) will flow if power would be connected to the tested appliance. Select YES or NO with Y or N key. Proceed with testing only if all safety measures have been taken. It is recommended to perform a thorough earth bond test on the PE of the appliance before proceeding with the test.
OmegaGTPlus / BetaGTPlus Test was skipped for safety! Instrument description Instrument skipped the required test because of a failed previous test. Instrument overheated! Measurement is aborted! Warning! More than 80 % of memory is occupated. Stored data should be downloaded to PC. Warning! Calibration has been expired. Temperature of internal components of the instrument reached their top limit. Measurement is prohibited until the internal temperature has reduced. Instrument memory is almost full.
OmegaGTPlus / BetaGTPlus Instrument description Warning! A high voltage is / will be present on the instrument output! (Insulation test voltage, flash test voltage (MI 3305 only) or mains voltage). The appliance under test should be switched on (to ensure that the complete circuit is tested). Connect the lead to be tested to the IEC plug. Connect current clamp adapter in this test. Test passed. Test failed. Some tests in the autotest sequence were skipped, but all performed tests passed. 2.
OmegaGTPlus / BetaGTPlus Main menu and test modes 3 Main menu and test modes The OmegaGTPlus / BetaGTPlus instrument has a user-friendly manipulation. Most of the actions can be done by using only few keys. Menu tree of the instrument is simple to understand and easy to operate. From the Main menu different instrument operation modes can be set: Ì Ì Ì Ì Ì Ì Ì Ì Ì VDE organizer (see 3.1), Custom autotest (see 3.2), Project autotests (see 3.3), Barcodes / RFID tags (see 3.4), Single test menu (see 3.
OmegaGTPlus / BetaGTPlus Ì Ì Main menu and test modes Test sequence can be executed, Test sequence can be stored as a configurable autotest. VDE organizer menu Keys: À/¿ ½/¾ F1 (VIEW) START ESC Select the item. Modify the item value. Opens view of all related tests and their parameters. Opens menu for performing tests according to organizer. Returns to Main menu. 3.1.1 VDE organizer – general VDE organizer is a configuration tool for performing VDE 0701/ VDE 0702 compatible appliance test sequences.
OmegaGTPlus / BetaGTPlus Main menu and test modes VDE Organizer flowchart With the VDE organizer any VDE 0701 or VDE 0702 compatible test sequence can be created. The sequences cover virtually any maintenance or periodic test, regardless of appliance type, safety class, supply cord length, fuse type, etc. All limits and tests comply with the currently valid VDE standards and regulations. In case of any changes, a firmware upgrade will be available at your distributor or from Metrel directly.
OmegaGTPlus / BetaGTPlus Main menu and test modes 3.1.2 VDE organizer operation Displayed data is organized as shown in VDE organizer flowchart. Only data relevant for selected standard and class is displayed, other data is hidden.
OmegaGTPlus / BetaGTPlus Main menu and test modes 3.1.3 Performing autotest sequences Once the any of autotest sequence is selected it can be performed. Keys: START ESC Starts of selected autosequence execution. Returns back main menu. Main principle is that after each successfully finished procedure in the autosequence, the next procedure is automatically prepared for execution and waits for operator to prepare new test setup (optional) and start the test.
OmegaGTPlus / BetaGTPlus Main menu and test modes Other measuring functions Measuring functions use common principle for access. Keys: START F2 (HELP) F4 (SKIP) F5 (END) ESC Starts the measurement. Displays test circuit for the measurement (test setup). Skip inspection to the next test function. Finish currently performing autotest. Returns back autotest entry menu. Key during measurement: STOP Stops the measurement.
OmegaGTPlus / BetaGTPlus Main menu and test modes Note: Ì The test probe for class II appliances could influence result! Insulation resistance measurement Note: Ì Use the test probe for class II appliances and class II parts of class I appliances! Insulation resistance measurement – Class II Note: Ì The test probe for class II appliances could influence result! Substitute leakage current measurement 21
OmegaGTPlus / BetaGTPlus Main menu and test modes Note: Ì Use the test probe for class II appliances and class II parts of class I appliances! Substitute leakage current measurement - Class II Notes: Ì If any of the earth bond resistance, insulation resistance, or substitute leakage current measurements have failed in selected autotest sequence, the differential current measurement is prohibited! Ì If leakage current measurement fails, test sequence will be finished.
OmegaGTPlus / BetaGTPlus Ì Main menu and test modes for measuring touch leakage current! The instrument automatically changes L/N polarity of connected appliance.
OmegaGTPlus / BetaGTPlus Main menu and test modes Note: Ì If any of earth bond resistance, insulation resistance, substitute leakage current, leakage current, touch leakage current or TRMS current clamp measurements have failed or were skipped in selected autotest sequence, the PRCD test is prohibited! RCD test 3.1.4 Example of creating a test sequence with VDE organizer A periodic test of an iron will be performed.
OmegaGTPlus / BetaGTPlus Main menu and test modes Selection of Earth bond limit value on base of known supply cable length: - Select appropriate length. Question if insulation test is applicable: - Confirm with yes. 5 Cord length (Earth bond) L: < 0.3Ω/<=5m 6 Insulation test applicable? YES 7 Insulation test Heating elements L: >0.3 MΩ 8 Insulation test Accessible cond. Parts L: >2.0 MΩ Leakage test method: Leakage Limit / Appliance type 10 General appliance L < 3.
OmegaGTPlus / BetaGTPlus Main menu and test modes Keys: À/¿ F1 (VIEW) F3 (DELETE) F5 (SEND) START ESC Select the specific test sequence. Opens view of selected test sequence with all related tests and their parameters. Removes selected test, see chapter 3.2.1 Deleting an existing custom test sequence. Sends autotest sequence to the RFID tag. Opens menu for performing tests according to organizer. See chapter 3.1.3 Performing autotest sequences. Returns to Main menu. 3.2.
OmegaGTPlus / BetaGTPlus F4 (EDIT) ¿/À START ESC Main menu and test modes Opens menu for modifying parameters of selected test function. Select test function in the sequence. Starts running the selected function. See chapter 3.1.3 Performing autotest sequences. Returns back to custom autotest menu. Modification of autotest sequence The key F4 (EDIT) as mentioned above enable to modify selected function. Modification parameters of selected test function Keys: À/¿ ½/¾ F1 (VIEW) ESC Select the parameter.
OmegaGTPlus / BetaGTPlus Main menu and test modes Save option Save as option Autotest custom save menus Keys: ½/¾ Alphanumeric SHIFT+ Alphanumeric Å F1 (SAVE) F2 (UNDO) ESC Select character in line. Enters character. Enters small letter or special character. Deletes character left to cursor. Confirms saving custom autotest sequence under entered name. Cancels saving. Returns back to custom autotest menu.
OmegaGTPlus / BetaGTPlus Ì Ì Ì Main menu and test modes Old test results do not occupy the instrument’s memory and can be temporarily uploaded only for the purpose of re-testing, Test results /appliance data can be moved / shared among different test instruments, Appliance data can be pre-entered on the computer and then sent to the instrument. Project Autotest menu example Keys: À/¿ ½/¾ Alphanumeric SHIFT+ Alphanumeric Å F1 (FIND) F2 (UNDO) F3 (TYPE) ESC Select the item. Select character in line.
OmegaGTPlus / BetaGTPlus Main menu and test modes Keys: À/¿ F1 (PgUp) F2 (PgDown) Select the appliance that should be retested. Starts execution of the autosequence for selected appliance (retest). See 3.1.3 Performing autotest sequences. Opens more detailed information of appliance results and data. Returns to Main menu. START ENTER ESC Appliance tickers Each appliance is marked with a ticker at the right end of the line.
OmegaGTPlus / BetaGTPlus Main menu and test modes Result of appliance (re)testing After finished testing the results can be: Ì Viewed, Ì Saved, Ì Sent to output device or PC. After finished autotest menu Keys: À/¿ ½/¾ alphanumeric SHIFT+ alphanumeric Å F1 (SAVE) F2 (UNDO) F3 (SEND) F4 (VIEW) ESC Select the device data. Select character in line. Enters character. Enters small letter or special character. Deletes character left to cursor.
OmegaGTPlus / BetaGTPlus Main menu and test modes 3.3.1 Comparison of results (evaluation of result trends) If an autotest sequence was performed from the Project Autotest menu an additional option TREND is offered in the View result menu after the autotest is finished but not saved. Trend of measured data Keys: À/¿ F1 (PgUp) F2 (PgDown) F5 (RESULT) ESC Scroll displayed items. Return to view the last results. Returns to Main menu.
OmegaGTPlus / BetaGTPlus Main menu and test modes Barcode / TAG menu Keys: À/¿ ENTER ESC Select the device. Opens menu for selected device. Returns to Main menu. 3.4.1 Working with RFID tags To use RFID system connect RFID reader/writer to DB-9 female BARCODE connector first. See RFID reader/writer user manual for more information.
OmegaGTPlus / BetaGTPlus Main menu and test modes Loading data from RFID tags to the instrument Select Tag test from Barcode / tag menu and press the ENTER key to confirm. The OmegaGTPlus / BetaGTPlus instrument is now ready to accept RFID tag data. RFID tag menu Once the data from RFID tag have been successfully received, the following menu is displayed: RFID tag menu Keys: À/¿ ENTER ESC Select the option. Opens menu for selected option. Returns to Barcode/tag menu.
OmegaGTPlus / BetaGTPlus Main menu and test modes How to view autotest results from RFID tag From the TAG menu select View results and press ENTER key to confirm. Refer to chapter 6.2.1 Results view. How to send an autotest sequence to RFID tag From the Autotest custom menu press the F5 (SEND) key. Selected autotest sequence is loaded to the RFID tag using RFID reader/writer.. See RFID reader/writer user manual for more information.
OmegaGTPlus / BetaGTPlus Main menu and test modes Reading autotest sequence using barcode From the Barcode/tag menu select Barcode test and press ENTER key. The following menu is displayed: Barcode test menu Keys: F1 (VIEW) START ½/¾ numeric keypad ESC Views selected autotest sequence. Refer to chapter 3.2.2 Viewing/modifying of existing custom autotest. Starts selected autotest sequence. Refer to chapter 3.1.3 Performing autotest sequence. Selects new autotest sequence. Returns to Barcode/tag menu.
OmegaGTPlus / BetaGTPlus Main menu and test modes 3.5 Single test menu In single test menu individual tests can be performed. Single test menu Keys: À/¿ ENTER ESC Select the test / measuring function: - , Resistance of bonding to protective earth connection. - , Insulation resistance test. - , Insulation resistance test with probe. - , Substitute leakage current test. - Substitute leakage current test with probe.
OmegaGTPlus / BetaGTPlus Main menu and test modes Keys: À/¿ ENTER ESC Select the item. Enters selected item menu. Returns to Main menu. Existing user names and appliance data can also be downloaded and edited using PC software and then uploaded to the instrument. Refer to the chapter 3.8 Data upload / download menu for detailed information. 3.6.1 Users submenu In this menu user names for up to 15 different users can be entered, edited and selected.
OmegaGTPlus / BetaGTPlus Main menu and test modes 3.6.2 Appliances submenu In this menu, default lists of appliance names (up to 100) can be edited. An alternative is to upload the names from PC software. Refer to the chapter 3.8 Data upload / download menu for detailed information. Edit Set appliances menu Keys: À/¿ F2 (PgUp) F3 (PgDown) F1 (EDIT) ENTER ESC Select the appliance. Opens menu for appliance name modification. Confirms selected appliance. Returns to users / appliance data menu.
OmegaGTPlus / BetaGTPlus ENTER ESC Main menu and test modes Confirms selected building. Returns to users / appliance data menu. Building rename See 3.6.1 Users submenu (User name modification). 3.6.4 Locations submenu In this menu default lists of location / room names (up to 100) can be edited. An alternative is to upload the names from the PC. The existing lists can also be downloaded to PC. Refer to the chapter 3.8 Data upload / download menu for detailed information.
OmegaGTPlus / BetaGTPlus Main menu and test modes Memory operation menu Parameter types are: [----, appliance, user, building, location, date]. Keys: À/¿ F1 (FIND) F2 (UNDO) F3 (TYPE) ESC Select the item. Starts searching stored results by applied parameters. Resets modified value to previous state. Modifies searching options by changing order or disabling (----) of parameter type. Returns to Main menu. See chapter 6 Data operations for more information. 3.
OmegaGTPlus / BetaGTPlus Main menu and test modes For detailed information on data uploading / downloading refer to help menus of PC software PATLink PRO. 3.9 Setup menu In this menu general instrument parameters can be set. Setup menu Keys: À/¿ ENTER ESC Select the setting. Opens menu for modification of selected setting. Returns to main menu.
OmegaGTPlus / BetaGTPlus Main menu and test modes Keys: ½/¾ À/¿ F1 (SAVE) F2 (UNDO) ESC Select the field to be changed. Select the setting. Saves new settings and returns to main settings menu. Recovers currently modified setting. Returns to main settings menu without changes. Notes: Ì Date is attached to the saved autotest results! Ì Date format is DD-MM-YYYY (day–month–year). Ì Date entry is checked for regularity and not accepted in case of irregular date! 3.9.
OmegaGTPlus / BetaGTPlus Main menu and test modes Keys: F1 (EDIT) ESC Starts editing of print heder. Returns to main settings menu. Editing mode for printing header Keys: À/¿ ½/¾ alphanumeric SHIFT+ alphanumeric Å F1 (SAVE) F2 (UNDO) ESC Select line. Character in line Enters character. Enters small letter or special character. Deletes character left to cursor. Saves new print header and returns to main settings menu. Recovers currently modified setting. Returns to main settings menu. 3.9.
OmegaGTPlus / BetaGTPlus Main menu and test modes Note: Ì Operator cannot change any instrument data! 3.9.5 Contrast adjustment Contrast can be adjusted with the contrast option. Contrast menu Keys: À/¿ F1 (SAVE) F2 (UNDO) ESC Adjust contrast. Saves new settings and returns to main settings menu. Recovers currently modified setting. Returns to main settings menu without changes. 3.9.
OmegaGTPlus / BetaGTPlus Main menu and test modes There are the following instrument settings: Item Appliance no.
OmegaGTPlus / BetaGTPlus Main menu and test modes Menu for setting initial values Keys: Y N, ESC Accepts initial settings. Returns to main setting menu without changes. Initial values: Parameters in single test mode User defined tests Autotest sequences Menus Baud rate Printer data transfer Reset all Clear all Added 24 pre-programmed autotests All set to initial structure. 115200 baud. Set to hardware flow control (DSR/DTR).
OmegaGTPlus / BetaGTPlus Main menu and test modes Communication settings Keys: À/¿ F1 (SAVE) ENTER ESC Select the option. Accepts selected option. Returns to communication settings menu. Note: Ì Only one port can be active at one time. 3.9.9 Password The instrument enables password protection for the following activities: Ì Ì Ì Ì Entering Edit user menu, Entering Edit measurements menu, Deleting stored results, Entering menu for Setting the instrument to initial values.
OmegaGTPlus / BetaGTPlus Main menu and test modes Opens menu for confirmation of new password. Returns to main settings menu without changes. ENTER ESC Confirmation request or change the existing password Confirmation or change existing password Keys: ½/¾ alphanumeric SHIFT+ alphanumeric Å ENTER ESC Select character in line. Enters character. Enters small letter or special character. Deletes character left to cursor. Confirms the password. Returns to main settings menu without changes.
OmegaGTPlus / BetaGTPlus Single test mode 4 Single test mode In single test mode individual tests can be performed. This is especially helpful when you suspect one or more faults in the tested appliance. Note: Ì The measurement results cannot be saved in this mode! Single test menu Entering the selected function will open its menu. Display organization in general can be seen in the figure below.
OmegaGTPlus / BetaGTPlus Measurements 5 Measurements 5.1 Earth bond resistance This test ensures that the connections between the protective conductor terminal in the mains plug of the appliance and accessible conductive parts of the appliance (metal housing) are satisfactory and of sufficiently low resistance. This test has to be performed on Class 1 (earthed) appliances. A high current is normally used to stress the connections under fault conditions.
OmegaGTPlus / BetaGTPlus Measurements Earth bond resistance measurement procedure Ì Ì Ì Ì Ì Ì Ì Ì Select the EARTH BOND function. Set output test current. Set limit value. Set measuring time. Connect device under test to the instrument. Connect earth bond clip into its socket on the instrument. Connect alligator of earth bond clip to accessible metal parts of the device under test (see figure above). Press the START key for measurement.
OmegaGTPlus / BetaGTPlus Measurements 5.2.1 Insulation resistance on class I appliances The instrument measures the insulation resistance between: Ì Ì Main test socket (L+N) and PE test terminals, LN and PE test sockets.
OmegaGTPlus / BetaGTPlus Measurements Insulation resistance measurement procedure for CLASS I Ì Ì Ì Ì Ì Ì Ì Ì Select the Insulation function. Set output test voltage. Set limit value. Set measuring time. Connect device under test to the instrument (see figures above). When appliance has isolated accessible metal parts, use S probe (earth bond clip), see 5.2.2 Insulation resistance on class 2 appliances.
OmegaGTPlus / BetaGTPlus Measurements Test circuits for CLASS II insulation resistance measurement Measurement of insulation resistance of class II appliance Measurement of insulation resistance of accessible isolated conductive parts of fixed installed appliances Insulation resistance measurement procedure for CLASS II Ì Ì Ì Ì Ì Ì Ì Ì Select the desired Insulation S. function. Set output test voltage. Set limit value. Set measuring time. Connect device under test to the instrument.
OmegaGTPlus / BetaGTPlus Measurements 5.3 Substitute leakage current Leakage currents between live conductors and isolated accessible metal parts (housing, screws, handles etc.) are checked with this test. Capacitive leakage paths are included in the result too. The current flowing at test voltage of 40 Vac is measured and scaled to the value at nominal mains supply voltage of 230 Vac.
OmegaGTPlus / BetaGTPlus Measurements Test circuits for CLASS I substitute leakage current measurement Measurement of substitute leakage current of class I appliance Measurement of substitute leakage current of fixed installed appliances of class I Substitute leakage current measurement procedure for CLASS I Ì Ì Ì Ì Ì Ì Ì Select the Sub leakage function. Set limit value. Set measuring time. Connect device under test to the instrument (see figures above).
OmegaGTPlus / BetaGTPlus Measurements Examples of substitute leakage current measurement results 5.3.2 Substitute leakage current on class II appliances The instrument measures leakage current between: Ì Ì Main test socket (L+N) and S test terminals, LN and S test sockets. Sub leakage S.
OmegaGTPlus / BetaGTPlus Measurements Measurement of substitute leakage of accessible isolated conductive parts of fixed installed appliances Substitute leakage current measurement procedure for CLASS II Ì Ì Ì Ì Ì Ì Select the Sub leakage S. function. Set limit value. Set measuring time. Connect device under test to the instrument (see figures above). For fixed equipment connect PE socket of the instrument to metallic enclosure of fixed equipment. Press the START key for measurement.
OmegaGTPlus / BetaGTPlus Measurements The instrument measures differential leakage current for both polarities of L and N. Differential current menu Test parameters for differential current measurement OUTPUT LIMIT TIME Test voltage [230 V] Maximum current [0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.50 mA, 3.50 mA, 4.00 mA, 4.50 mA, 5.00 mA, 5.50 mA, 6.00 mA, 7.00 mA, 8.00 mA, 9.
OmegaGTPlus / BetaGTPlus Measurements Notes: Ì The appliance is connected to the mains voltage. Make sure that the appliance with moving parts is safely mounted or protected to prevent possible danger to operator or damage to the appliance! Ì Consider any displayed warning before starting measurement! Ì The instrument automatically changes L/N polarity of connected appliance. 5.5 Touch leakage current This test determines the current that would flow if a person touches the appliance.
OmegaGTPlus / BetaGTPlus Measurements Measurement of touch leakage current on a fixed installed appliance Touch leakage current measurement procedure Ì Ì Ì Ì Ì Select the Touch leakage function. Set limit value. Set measuring time. Connect device under test to the instrument (see figure above). Press the START key for measurement. Examples of touch leakage current measurement results (slika) Notes: Ì The appliance is connected to the mains voltage.
OmegaGTPlus / BetaGTPlus Measurements Polarity test menu Test circuit for polarity test Polarity test of IEC cord Polarity test procedure Ì Ì Ì Select the Polarity test function. Connect tested cable to the instrument (see figure above). Press the START key for measurement.
OmegaGTPlus / BetaGTPlus Measurements 5.7 Clamp current measurement This function enables measuring of AC currents in a wide range from 1 mA up to 25 A. Typical applications are: Ì Ì Ì Measuring leakage currents through PE conductor in permanently installed appliances, Measuring load currents through L or N conductor in permanently installed appliances, Measuring residual current in the group of conductors, etc.
OmegaGTPlus / BetaGTPlus Measurements Test circuit for clamp current measurement Connecting current clamp to the instrument Clamp current measurement procedure Ì Ì Ì Ì Ì Ì Select the Clamp current function. Set limit value. Set measuring time. Connect the current clamp to the instrument (see figure above). Put the current clamp around wire(s) that has to be measured. Press the START key for measurement.
OmegaGTPlus / BetaGTPlus Measurements 5.8 RCD test The purpose of this test is to ensure proper operation of residual current devices built-in in appliances and portable residual current devices. The following measurements and tests can be performed: Ì Trip-out time, and Ì RCD autotest.
OmegaGTPlus / BetaGTPlus Measurements Circuits for testing RCD Testing of portable RCD (PRCD) 5.8.1 Single RCD trip-out time test Trip-out time measurement verifies the sensitivity of an RCD at different residual currents. Trip-out time measurement procedure Ì Ì Ì Ì Ì Ì Ì Select the RCD test function. Select Single test mode. Set test parameters (if necessary). Connect tested device to the test socket of the instrument (see figures above).
OmegaGTPlus / BetaGTPlus Measurements false parameter is noticed during the RCD autotest, then individual parameter test has to be used for further investigation. RCD auto test menu RCD autotest procedure RCD Autotest steps Ì Select the RCD test function. Ì Set Auto test mode. Ì Select rated IΔN (if necessary). Ì Connect tested device to the test socket of the instrument (see figures above). Ì Connect test lead from PRCD socket of the instrument to the secondary side of tested PRCD, terminal L.
OmegaGTPlus / BetaGTPlus Measurements Step 3 Step 4 Step 5 and Step 6 Individual steps in RCD autotest The test passes if the RCD: Ì Ì Does not trip out at ½×IΔN tests, Trips inside predefined time limits at IΔN, and 5×IΔN tests. Notes: Ì Consider any displayed warning before starting measurement! Ì For appliances with integrated RCD the housing must be opened to access the RCD’s L output terminal. Ì Mains voltage is applied to the device under test.
OmegaGTPlus / BetaGTPlus Measurements Test parameters for Functional test OUTPUT TIME System voltage [230 V] Measuring time [5 s, 10 s, 30 s, 60 s, 120 s, 180 s] Circuit for functional test Functional test Functional test procedure Ì Ì Ì Ì Select the Functional test function. Set measuring time. Connect tested appliance to the instrument (see figure above). Press the START key for measurement. Example of apparent power measurement result Notes: Ì The appliance is connected to the mains voltage.
OmegaGTPlus / BetaGTPlus Measurements Flash test menu Test parameters for flash test OUTPUT LIMIT TIME Test voltage [1500 V, 3000 V] Maximum current [1.00 mA, 1.50 mA, 2.00 mA, 2.25 mA] Measuring time [5 s, 10 s, 30 s, 60 s, 120 s, 180 s] Test circuits for flash test For Class I appliances, a test voltage of 1.5 kVAC is applied between main test socket (L+N) and PE test terminals.
OmegaGTPlus / BetaGTPlus Measurements Flash test procedure for CLASS I: Ì Ì Ì Ì Ì Ì Select the Flash test function. Set output test voltage. Set limit value. Set measuring time. Connect device under test to the instrument (see figures above). Press the START key for measurement. Flash test procedure for CLASS II: Ì Ì Ì Ì Ì Ì Ì Ì Select the Flash test function. Set output test voltage. Set limit value. Set measuring time. Connect device under test to the instrument (see figures above).
OmegaGTPlus / BetaGTPlus Data operations 6 Data operations After the autotest sequence is completed, measurement results can be: Ì Ì Ì Ì Ì Viewed, Saved to the flash memory of the instrument, Send to PC, Test report can be send to print out on serial printer, Appliance label can be printed out. 6.1 Saving autotest results After the autotest sequence is completed, Save results menu is displayed. Important data can be added to the autotest results in this menu.
OmegaGTPlus / BetaGTPlus Data operations Building data for items Up to 10 alphanumeric or special characters can be entered. Appliance no. Appliance number can also be read with a barcode reader. Test site, Up to 15 alpha-numeric or special characters can be entered. Names can also be selected from the list of 100 predefined names Location, Appliance name for each parameter. 2 numeric characters can be entered. Retest period Up to 20 alpha-numeric or special characters can be entered.
OmegaGTPlus / BetaGTPlus Data operations 6.2.1 Results view For each autotest result can be observed: Ì Ì Ì All functions applied, Test parameters for each and Results with Pass/Fail indication. View results menu Keys: ¿, F1 (PgUp) À, F2 (PgDown) F5 (MORE) ESC Opens previous page of test results. Opens next page of test results. Switches between detailed results view and complete list of appliance data. Returns memory recall menu. 6.2.
OmegaGTPlus / BetaGTPlus Data operations Keys to confirm deleting: Confirms deleting in deletes selected option. Y N Returns to memory recall menu without deleting. ESC 6.2.3 Downloading and printing results The instrument offers the following possibilities to send selected result or selection to: PC (to store and later operations on stored results), serial printer for fast report printing, label printer, and RFID tag.
OmegaGTPlus / BetaGTPlus Data operations Serial printer Send to serial printer menu Keys for transfer to serial printer: Opens menu for interface selection (back site printer or F1 (COM) PC/Printer connector). Opens menu for selection of data transfer control option. F2 (Control) Interface option Data transfer control Serial printer settings Keys: À/¿ F1 (SAVE) ENTER ESC Select the option. Accepts selected option. Returns to Send results menu.
OmegaGTPlus / BetaGTPlus Data operations RFID tag Send to tag menu Keys: ENTER ESC Send data to the RFID tag. Returns to Send results menu. Notes: Ì When working with a serial printers, baud rate is set by default to 9600 bps.
OmegaGTPlus / BetaGTPlus Maintenance 7 Maintenance 7.1 Periodic calibration It is essential that all measuring instruments are regularly calibrated in order for the technical specification listed in this manual to be guaranteed. We recommend an annual calibration. The calibration should be done by an authorized technical person only. 7.2 Fuses There are two fuses available from front panel: F1 = F2 = T 16 A / 250 V (32 × 6,3 mm): intended for instrument protection.
OmegaGTPlus / BetaGTPlus Technical specifications 8 Technical specifications 8.1 Earth bond resistance Earth bond resistance readout (4 A, 10 A, 25 A) Range 0.00 Ω ÷ 1.99 Ω 2.00 Ω ÷ 19.99 Ω Resolution 0.01 Ω 0.01 Ω Accuracy ±(5 % of reading + 3 digits) ±10 % Earth bond resistance readout (200 mA) Range 0.00 Ω ÷ 1.99 Ω 2.00 Ω ÷ 9.99 Ω 10.0 Ω ÷ 19.9 Ω Resolution 0.01 Ω 0.01 Ω 0.1 Ω Accuracy ±(5 % of reading + 3 digits) ±(5 % of reading + 5 digits) ±(5 % of reading + 5 digits) Test currents............
OmegaGTPlus / BetaGTPlus Technical specifications Class II Insulation resistance readout Range 0.000 MΩ ÷ 0.500 MΩ 0.501 MΩ ÷ 1.999 MΩ 2.00 MΩ ÷ 19.99 MΩ 20.0 MΩ ÷ 199.9 MΩ Resolution 0.001 MΩ 0.001 MΩ 0.01 MΩ 0.1 MΩ Accuracy ±(10 % of reading + 5 digits) ±(5 % of reading + 3 digits) Nominal voltages ...... 250 V DC, 500 V DC (- 0 %, + 10 %) Measuring current..... min. 1 mA at 250 kΩ (250 V), 500 kΩ (500 V) Short circuit current... max. 2.0 mA Pass levels ............... 0.10 MΩ, 0.25 MΩ, 0.50 MΩ, 1.
OmegaGTPlus / BetaGTPlus Technical specifications 8.4 Differetial current Differential leakage current readout Range 0.00 mA ÷ 9.99 mA Resolution 0.01 mA Accuracy ±(5 % of reading + 5 digits) Pass levels ............... 0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.50 mA, 3.50 mA, 4.00 mA, 4.50 mA, 5.00 mA, 5.50 mA, 6.00 mA, 7.00 mA, 8.00 mA, 9.00 mA, none Test duration............. 5 s, 10 s, 30 s, 60 s, 120 s, 180 s Output....................... test socket Polarity L-N ...............
OmegaGTPlus / BetaGTPlus Technical specifications Test duration............. 5 s, 10 s, 30 s, 60 s, 120 s, 180 s Input current clamp terminals 8.8 Trip-out time of portable RCD Portable RCD trip-out time readout Range 0 ms ÷ 1999 ms(½×IΔN) 0 ms ÷ 300 ms (IΔN) 0 ms ÷ 40 ms (5×IΔN) Resolution 1 ms 1 ms 1 ms Accuracy ±3 ms Test currents (IΔN) ............. 10 mA, 15 mA, 30 mA Test current multipliers ..... ½×IΔN, IΔN, 5×IΔN Start angle ........................ 0°, 180°, both Test modes ..................
OmegaGTPlus / BetaGTPlus Technical specifications 8.11 General data Rated supply voltage .........................230 V AC (±10 %, 50 or 60 Hz) Max. power consumption...................150 VA (without tested appliance) Maximum appliance current...............16 A Overvoltage category.........................Cat II / 300 V Protection classification .....................I Pollution degree.................................2 Degree of protection ..........................IP 50 (closed and locked cover) Case ..
OmegaGTPlus / BetaGTPlus Instrument set and accessories 9 Instrument set and accessories Standard set of the instrument Ì Ì Ì Ì Ì Ì Ì Ì Ì Ì Ì Ì Ì Ì Instrument MI 3305 – OmegaPATPlus / MI 3304 – BetaPATPlus Additional bag for accessories inside carrying case Test probe (earth bond clip) Flash test probe (MI 3305 only) PRCD test probe Test lead (1.5 m, green) Test lead (1.
OmegaGTPlus / BetaGTPlus Appendix A Appendix A 24 pre-programmed autotest sequences No. Name 1 702_Kl_1_Iso 2 702_Kl1_Iso_BLT 3 702_Kl_1_Ia 4 702_Kl_1_Ia_BLT 5 702_Kl_2_Iso 6 702_Kl_2_Ibs 7 702_Kl_1_IsoIa 8 702Kl1_IsoIaBLT 9 702_Kl_2_IsoIbs 10 702_Kl_2 11 701_Kl_1_Iso 12 701_Kl1_Iso_BLT Description Testing according to VDE 702. Class 1 appliance. Insulation resistance and substitute leakage current measurements are selected. Testing according to VDE 702.
OmegaGTPlus / BetaGTPlus Appendix A 24 pre-programmed autotest sequences (cont’d) No. Name 13 701_Kl_1_Ia 14 701_Kl_1_Ia_BLT 15 701_Kl_2_Iso 16 701_Kl_2_Ibs 17 701_Kl_1_IsoIa 18 701Kl1_IsoIaBLT 19 701_Kl2_Iso_Ibs 20 701_Kl_2 21 701_Kl_3_Iso 22 702_Kl_3_Iso 23 701_Kl_3 24 702_Kl_3 Description Testing according to VDE 701. Class 1 appliance. Differential leakage current measurement is selected. Testing according to VDE 701.
OmegaGTPlus / BetaGTPlus Appendix A 24 pre-programmed autotest sequences table Autotest shortcut code Visual test Output Earth bond Limit Time Output Insulation Limit Time Output Insulation Limit (probe) Time Output Sub leakage Limit Time Output Sub leakage Limit (probe) Time Output Differential Limit leakage Time Output Touch Limit leakage Time Output Power Limit Time Output TRMS clamp Limit current Time Polarity test 01 702_Kl_1_Is o ; 200 mA 0.30 Ω 5s 500 V 1.00 MΩ 5s : : : 40 V 3.
OmegaGTPlus / BetaGTPlus Appendix A 24 pre-programmed autotest sequences table (cont’d) Autotest shortcut code Visual test Output Limit Time Output Insulation Limit Time Output Insulation Limit (probe) Time Output Sub leakage Limit Time Output Sub leakage Limit (probe) Time Output Differential Limit leakage Time Output Touch Limit leakage Time Output Power Limit Time Output TRMS clamp Limit current Time Polarity test Earth bond 05 06 07 08 702_Kl_2_Iso 702_Kl_2_Ibs 702_Kl_1_IsoIa 702_Kl1_IsoIaBLT
OmegaGTPlus / BetaGTPlus Appendix A 24 pre-programmed autotest sequences table (cont’d) Autotest shortcut code Visual test Output Limit Time Output Insulation Limit Time Output Insulation Limit (probe) Time Output Sub leakage Limit Time Output Sub leakage Limit (probe) Time Output Differential Limit leakage Time Output Touch leakage Limit Time Output Power Limit Time Output TRMS clamp Limit current Time Polarity test Earth bond 09 702_Kl_2_IsoIbs ; : : : : : : 500 V 2.00 MΩ 5s : : : : : : : : : 230 V 0.
OmegaGTPlus / BetaGTPlus Appendix A 24 pre-programmed autotest sequences table (cont’d) Autotest shortcut code Visual test Output Limit Time Output Insulation Limit Time Output Insulation Limit (probe) Time Output Sub leakage Limit Time Output Sub leakage Limit (probe) Time Output Differential Limit leakage Time Output Touch Limit leakage Time Output Power Limit Time Output TRMS clamp Limit current Time Polarity test Earth bond 13 701_Kl_1_Ia ; 200 mA 0.30 Ω 5s : : : : : : : : : : : : 230 V 3.
OmegaGTPlus / BetaGTPlus Appendix A 24 pre-programmed autotest sequences table (cont’d) Autotest shortcut code Visual test Output Earth bond Limit Time Output Insulation Limit Time Output Insulation Limit (probe) Time Output Sub leakage Limit Time Output Sub leakage Limit (probe) Time Output Differential Limit leakage Time Output Touch Limit leakage Time Output Power Limit Time Output TRMS clamp Limit current Time Polarity test 17 701_Kl1_IsoI a ; 200 mA 0.30 Ω 5s 500 V 1.
OmegaGTPlus / BetaGTPlus Appendix A 24 pre-programmed autotest sequences table (cont’d) Autotest shortcut code Visual test Output Limit Time Output Insulation Limit Time Output Insulation Limit (probe) Time Output Sub leakage Limit Time Output Sub leakage Limit (probe) Time Output Differential Limit leakage Time Output Touch Limit leakage Time Output Power Limit Time Output TRMS current Limit clamp Time Polarity test Earth bond 21 701_Kl_3_Iso ; : : : : : : 500 V 0.
OmegaGTPlus / BetaGTPlus Appendix A METREL GmbH VDE tester test type card Code Autotest sequence name and descriptions 01 Testing according to VDE 702. Class 1 appliance. Insulation resistance and substitute leakage current measurements are applicable. 02 03 04 05 06 07 08 702_Kl _1_Iso Limits Earth bond: 0.30 Ω Insulation: 1.00 MΩ Sub leakage: 3.50mA Earth bond: 0.30 Ω Insulation: 1.00 MΩ Insulation - S: 2.00 MΩ Sub leakage: 3.50 mA Sub leakage - S: 0.
OmegaGTPlus / BetaGTPlus Appendix A METREL GmbH VDE tester test type card (cont'd) Code 09 10 11 12 13 14 15 16 Autotest sequence name and descriptions 702_Kl _2_IsoI bs Testing according to VDE 702. Class 2 appliance with isolated accessible conductive parts. Prüfungen für Isolation und Berührungsstrom werden eingestellt. 702_Kl _2 Testing according to VDE 702. Class 2 appliance without any isolated accessible conductive parts. Limits Insulation - S: 2.00 MΩ Touch leakage: 0.
OmegaGTPlus / BetaGTPlus Appendix A METREL GmbH VDE tester test type card (cont'd) Code 17 18 19 20 21 22 23 24 Autotest sequence name and descriptions 701_Kl_1 _IsoIa 701Kl1_I soIaBLT 701_Kl2_ Iso_Ibs 701_Kl_2 701_Kl_3 _Iso 702_Kl_3 _Iso Limits Testing according to VDE 701. Class 1 appliance. Prüfungen für Isolation und Differenzstrom werden eingestellt. Earth bond: 0.30 Ω Insulation: 1.00 MΩ Leakage: 3.50 mA Testing according to VDE 701.
OmegaGTPlus / BetaGTPlus Appendix B Appendix B The instrument OmegaGTPlus / BetaGTPlus supports two barcode formats when printing appliance labels. Autotest shortcut code and appliance number Autotest shortcut codes are represented as a two digit code. These autotest codes can also be represented by the barcode. Using the barcode reader, the instruments can accept autotest shortcut code from barcode label. 01 Autotest shortcut code Also appliance number can be read from barcode label.
OmegaGTPlus / BetaGTPlus 01 $ 4455821981 Appendix B Autotest shortcut code Separator Appliance number Refer to chapter 3.9.6 Instrument settings for barcode system selection. Notes: Ì Special character »$« between autotest shortcut code and appliance name (ID number) is used to distinguish shortcut code from appliance name. Ì Only appliance ID is printed out on the 2nd appliance label (power supply cord label).
OmegaGTPlus / BetaGTPlus Appendix B 99
OmegaGTPlus / BetaGTPlus Appendix B 100