Operation Manual
OmegaGTPlus / BetaGTPlus Main menu and test modes
25
5
Cord length (Earth bond)
L: < 0.3Ω/<=5m
Selection of Earth bond limit value on base
of known supply cable length:
- Select appropriate length.
6 Insulation test applicable? YES
Question if insulation test is applicable:
- Confirm with yes.
Insulation and substitute leakage
measurements will be included in the test
sequence with this confirmation.
7
Insulation test
Heating elements L: >0.3 MΩ
Classification of appliance:
- Classify the iron as a standard appliance.
8
Insulation test
Accessible cond. Parts L: >2.0 MΩ
Information that insulation resistance
measurement of Class 2 parts will be
included in the (VDE702 / Class1 / with
isolated conductive parts) test procedure.
9
Leakage test method:
Leakage
Selection of leakage current test method:
- Select substitute leakage measurement.
10
Limit / Appliance type
General appliance L < 3.5mA
Classify the iron as an standard appliance
with power <3.5 kW.
11
Touch leakage method: Substitute
leakage
Limit < 0.5mA
Information that substitute leakage current
measurement class 2 parts will be included
in the (VDE702 / Class1 / with isolated
conductive parts) test procedure.
3.2 Custom autotests menu
The menu contains list of custom prepared autosequences. The sequence can be
performed or can be modified / adapted for special appliance and stored or sent to the
PC for further purposes (as well test reports can be printed out, etc).
24 pre-programmed autotest sequences are added to the list by default.
Up to 50 custom autotest sequences can be pre-programmed in this autotest mode.
An alternative is to pre-program the sequences in the PC software PATLink LITE and
upload them to the instrument. Refer to chapter 3.8 Upload test data for more
information.
The initial autotest sequences can be restored to default settings by selecting Original
settings in Setup menu.
Custom autotest menu