Operation Manual
OmegaGTPlus / BetaGTPlus Main menu and test modes
46
There are the following instrument settings:
Item Possible settings
Appliance no. [increment, replicate, blank]
Test site [replicate, blank]
Location [replicate, blank]
User [replicate, blank]
Appliance name [replicate, blank]
Retest period [replicate, blank]
Repairing code [replicate, blank]
Comment [replicate, blank]
Barcode system [single, double]
Meaning of possible settings are:
blank Appropriate field is set to blank in Save results menu.
replicate Last entered string is used in an appropriate field when new autotest
sequence is finished.
increment Appliance number is automatically incremented when new autotest
sequence is finished.
Modification of selected setting
Keys:
À / ¿
Modify setting of selected item.
F1 (SAVE)
Saves setting of selected item,
F2 (UNDO)
Recovers currently modified setting.
ESC
Returns to main settings menu.
The instrument supports special label printers (Printek MtP, Intermec EasyCoder C4).
When printing barcodes to the special printer, two different options can be set in
Barcode systems submenu. If single option is selected, only appliance name is
printed out as a barcode on appliance barcode label. When double is selected, both
autotest shortcut code and appliance name are printed out as a barcode on appliance
barcode label.
3.9.7 Setting the instrument to initial values
The Original settings menu sets some parameters of the instrument into their initial
values.