5.8 RCD test
The purpose of this test is to ensure proper operation of residual current devices built-in
in appliances and portable residual current devices.
The following measurements and tests can be performed:
Ì Trip-out time, and
Ì RCD autotest.
RCD single test menu RCD autotest menu
Test parameters for RCD test
I
ΔN
Rated RCD residual current [10 mA, 15 mA, 30 mA]
MODE Type of RCD test [single, auto]
If MODE = single
Multi
Actual test current I
ΔN
[x ½, x 1, x 5]
Phase
Starting angle [0°, 180°, (0°,180°)]
RCD test current starting polarities
Trip-out time limits
Trip-out times according to EN 61008 / EN 61009:
½×I
ΔN
*)
I
ΔN
5×I
ΔN
General RCDs
(non-delayed)
t
Δ
> 300 ms t
Δ
< 300 ms t
Δ
< 40 ms
*)
Minimum test period for current of ½×I
ΔN
, RCD shall not trip-out.
Maximum test times related to selected test current for general (non-delayed) RCD
Standard
½×I
ΔN
I
ΔN
5×I
ΔN
EN 61008 / EN 61009 400 ms 400 ms 40 ms
The instrument is intended for testing of General (non-delayed) RCDs, which are suited
for:
Ì Alternating residual current (AC type, marked with symbol),
PRCD socket of the instrument operates as current sink for fault current which is
sourced on L entry of tested appliance.