Operation Manual

OmegaGTPlus / BetaGTPlus Measurements
71
Flash test menu
Test parameters for flash test
OUTPUT Test voltage [1500 V, 3000 V]
LIMIT Maximum current [1.00 mA, 1.50 mA, 2.00 mA, 2.25 mA]
TIME Measuring time [5 s, 10 s, 30 s, 60 s, 120 s, 180 s]
Test circuits for flash test
For Class I appliances, a test voltage of 1.5 kV
AC
is applied between main test socket
(L+N) and PE test terminals.
Flash test on Class I appliance
For Class II appliances, a test voltage of 3 kV
AC
is applied between main test socket
(L+N) and FLASH test terminals.
Flash test on Class II appliance