Operation Manual

OmegaGTPlus / BetaGTPlus Measurements
72
Flash test procedure for CLASS I:
Ì Select the Flash test function.
Ì Set output test voltage.
Ì Set limit value.
Ì Set measuring time.
Ì Connect device under test to the instrument (see figures above).
Ì Press the START key for measurement.
Flash test procedure for CLASS II:
Ì Select the Flash test function.
Ì Set output test voltage.
Ì Set limit value.
Ì Set measuring time.
Ì Connect device under test to the instrument (see figures above).
Ì Connect flash probe into its socket on the instrument.
Ì Connect flash probe to the device under test (see figures above).
Ì Press the START key for measurement.
Examples of flash test results
Notes:
Ì Consider any warning on the display before starting measurement.
Ì Do not touch the equipment under test or the test leads during the test – a
dangerous voltage will be present.
Ì Appliance ON/OFF switches must be switched on.
Ì Before applying the test, check that the components in the appliance can
withstand the test voltage without any damage. IT appliances are especially
sensitive.