24 pre-programmed autotest sequences (cont’d)
No. Name Description
13
701_Kl_1_Ia
Testing according to VDE 701.
Class 1 appliance.
Differential leakage current measurement is selected.
14
701_Kl_1_Ia_BLT
Testing according to VDE 701.
Class 1 appliance with isolated accessible conductive parts.
Differential leakage current and touch leakage current
measurements are selected.
15
701_Kl_2_Iso
Testing according to VDE 701.
Class 2 appliance with isolated accessible conductive parts.
Insulation resistance and substitute leakage current
measurements are selected.
16
701_Kl_2_Ibs
Testing according to VDE 701.
Class 1 appliance.
Touch leakage current measurement is selected.
17
701_Kl_1_IsoIa
Testing according to VDE 701.
Class 1 appliance.
Insulation resistance and differential leakage current
measurements are selected.
18
701Kl1_IsoIaBLT
Testing according to VDE 701.
Class 1 appliance with isolated accessible conductive parts.
Insulation resistance, differential leakage current and touch
leakage current measurements are selected.
19
701_Kl2_Iso_Ibs
Testing according to VDE 701.
Class 2 appliance with isolated accessible conductive parts.
Insulation resistance and touch leakage current measurements
are selected.
20
701_Kl_2
Testing according to VDE 701.
Class 2 appliance without any isolated accessible conductive
parts.
21
701_Kl_3_Iso
Testing according to VDE 701.
Class 3 appliance with isolated accessible conductive parts.
22
702_Kl_3_Iso
Testing according to VDE 702.
Class 3 appliance with isolated accessible conductive parts.
23
701_Kl_3
Testing according to VDE 701.
Class 3 appliance without any isolated accessible conductive
parts.
24
702_Kl_3
Testing according to VDE 702.
Class 3 appliance without any isolated accessible conductive
parts.